REMOVE blister pack and CLIP filter to the
C-Clips so the flow symbol on filter is towards
the faucet.
CONNECT the water valve flex line to the inlet
of filter.
CONNECT the faucet stem flex line to the outlet
of filter.
OPEN water supply valves and faucet valves
to allow water to run through the filter for 3-5
minutes to flush out the disinfectant and purge
air out of the system.
USE residual test strips on running water to
ensure residual disinfectant levels do not exceed
4ppm.
CLOSE the faucet valves and CHECK for
any leaks. Tighten any fittings if necessary. If
included, PRESS ON the DSU Cover onto the
filter.
*It is recomended to install the DSU Cover to
shield the filter from the external environment.
CONNECT
CONNECT
FLOW
Installation Guide
Part #:
70-0285
(See DSU-H Instructions for Use)
DSU-H Sink (continued)
AVOID contact between the filter and the cleaning or sanitizing agents.
DO NOT remove blister pack from filter until indicated to protect from
contamination.