Questions? Call Neogen at 800-234-5333 or 517-372-9200
7
Setting the Pass, Marginal and Fail thresholds
The AccuPoint Advanced HC reader comes programmed with default threshold
settings of 0–149 as Pass, 150–299 as Marginal and 300 and higher as Fail. These
settings can be changed through the Data Manager program and downloaded to
the reader.
BOOT MENU
With HC firmware version 3.1 and higher, additional controls have been added to a
Boot Menu. To access this menu, (with the reader off) press and hold the down arrow
while pressing the power button. Contained within the boot menu are 4 options:
1.
Time and Date Lock:
when on, will prevent the time and date from being changed
on the main user screen.
2.
Power Saving Mode:
when on, will dim the display after 3 minutes of inactivity
and turn off display after 15 minutes. To recover the reader from sleep mode,
simply press the OK button and the display will return.
3.
Factory Reset:
when selected, will return the reader to its default configuration
and will remove all stored data, test plans, housekeepers, etc.
4.
Brightness:
after pressing the OK button, a slider appears on the display screen.
Repeatedly press the navigation keys to the left to dim and to the right to brighten.
After changes have been made, press the OK button to exit.
5.
Exit:
to exit the boot menu, press the OK button.
RECORD COUNT AND TEST RECALL
The reader keeps track of the number of tests that have been taken. The number of
records currently stored in the reader’s memory is displayed at the User ID screen
when the reader is powered on. The number of tests that have been taken since the
reader was powered on is displayed in the bottom right corner of the display.
Test results can be reviewed by navigating to the Record Count field. Press the OK
button and enter the field. Moving the navigation keys up or down within the field
causes the results to be displayed for the tests that have been taken since the reader
was powered on. To exit, press the left arrow.