AccuPoint 2 User’s Guide
10
SETTING THE TIME AND DATE
Toggle to the time or date field and depress the toggle button. Move the toggle button
to the appropriate digit and increment by moving up or down. The time display can be
set to a.m./p.m. or 24 hour mode. Setting the time field to 24 hour mode causes the
date field to display in the format DD/MM/YY.
POWERING THE ACCUPOINT 2 SYSTEM
The AccuPoint 2 instrument can only be operated from its battery power. The USB
power capability has been disabled to protect the integrity of the electrical circuit.
While the instrument will appear to power on when the A/C power is connected, it
can only operate in charging mode. No readings can take place when the instrument
is charging.
RECHARGING THE BATTERIES
The AccuPoint 2 instrument is supplied with 4 AA rechargeable nickel metal-hydride
(NiMH) batteries, which typically provide up to 16 hours of continuous operation before
requiring recharging. For optimal battery life, the instrument should be operated in
temperatures of 10–35°C, and the batteries should only be recharged when the
battery charge icon appears near depletion.
The AccuPoint 2 instrument can only be recharged through the supplied A/C power
adapter. Full charge takes approximately 90 minutes. After inserting the A/C power
cord, the instrument will display a green battery with the appropriate charge state.
This icon will change and fill in completely as the batteries become fully charged.
The icon will change to black and the instrument will power off when the charge is
complete.
The battery charge icon indicates the charge state of the instrument.
The rechargeable NiMH batteries supplied by Neogen have a normal lifetime of
approximately 1,000 recharging cycles, and are the only batteries warranted for use
in the instrument.
NOTES:
For best results, use only rechargeable batteries supplied by Neogen; however,
quality standard alkaline batteries can be used to power the instrument. If using
alkaline batteries, DO NOT attempt to charge them. Attempting to charge standard
alkaline batteries may damage the instrument.
The instrument’s data readings, and marginal and fail limits, will remain stored even
when the batteries have discharged.
Fully charged
Fully discharged
Charging