SRC-MAX Operation with a Pulse Oximeter or Monitor
12
The technician must perform the test setup procedure before
performing tests 1 through 4. The following is a brief
description of each test:
•
Test Setup
— This procedure establishes the
baseline for all the other tests. The Test Setup
procedure must be performed before performing any
or all of the SRC-MAX tests.
•
Test 1: BPM
— This procedure simulates an
O
XI
M
AX
sensor attached to a patient indicating 60
BPM and 200 BPM. The test setup procedure sets up
Test 1 for 60 BPM.
•
Test 2: SpO
2
— This procedure simulates an
O
XI
M
AX
sensor attached to a patient ,indicating 75
percent blood oxygen saturation and 90 percent
blood oxygen saturation. The test setup procedure
sets up Test 2 for 75 percent blood oxygen saturation.
•
Test 3: Modulation
— This procedure simulates an
O
XI
M
AX
sensor attached to a patient indicating low
and high pulse strength. The test setup procedure sets
up Test 3 for low pulse strength.
•
Test 4: Light
— This procedure simulates an
O
XI
M
AX
sensor attached to a patient indicating low
and high light level passing through the patient at the
sensor site. The test setup procedure sets up Test 4
for low light level.
Test Setup
Note:
Refer to the applicable pulse oximeter or monitor
operator’s manual for the pulse oximeter or monitor
operation and test procedures.
1. Locate the Sp0
2
Sensor Port on the front panel of the
pulse oximeter or monitor.
2. Connect a DOC-10 pulse oximetry cable to the pulse
oximeter’s Sp0
2
sensor port or the applicable OEM cable
to the monitor’s Sp0
2
sensor port. See Figure 7.
Содержание OxiMax SRC-MAX
Страница 1: ...SRC MAX Pulse Oximetry Functional Tester Technical Manual...
Страница 6: ...Contents SRC MAX iv Ta b l e s Table 1 Troubleshooting 32...
Страница 10: ...Blank Page...
Страница 16: ...SRC MAX Components and Accessories 10 C Lock Cable See Figure 6 Figure 6 C Lock Cable...
Страница 26: ...Blank Page...
Страница 40: ...Blank Page...
Страница 42: ...Blank Page...
Страница 48: ...Blank Page...
Страница 51: ...Index SRC MAX 45 Three Lead ECG cable 9 Transport Relative Humidity 40 Temperature 40 Troubleshooting 31...
Страница 52: ...Blank Page...
Страница 53: ...Blank Page...