Troubleshooting
O
XI
M
AX
N-65 Service Manual
41
• Replacing the front case does not fix the
problem.
Replace the SpO
2
PCB. See
SpO2 PCB or User Interface
PCB Replacement
on page 50.
• Replacing the SpO
2
PCB does not fix the
problem.
Replace the User Interface PCB. See
SpO2 PCB or User
Interface PCB Replacement
on page 50. If this does not fix
the problem, return the monitor to Nellcor.
One or more keys on the front panel keypad does not work.
• The flex circuit between the front panel
and the SpO
2
PCB is damaged, or not
making a good connection.
Caution: Unlock J4 before attempting to
remove and reinsert the flex cable.
Inspect the flex cable for damage. Inspect the connection to
J4. If there is damage to the flex cable, replace the front
case. Otherwise, remove the cable from J4 and reconnect it.
Reassemble the monitor and retest. Refer to
Separating the
Front and Back Cases
on page 48 and
Front Case Assembly
Replacement
on page 54.
• The front panel keypad is defective.
Caution: Unlock J4 before attempting to
remove the flex circuit cable.
Replace the front case. Refer to
Front Case Assembly
Replacement
on page 54.
• Replacing the front case does not fix the
problem.
Replace the SpO
2
PCB.
SpO2 PCB or User Interface PCB
Replacement
on page 50.
• Replacing the SpO
2
PCB does not fix the
problem.
Replace the User Interface PCB. See
SpO2 PCB or User
Interface PCB Replacement
on page 50. If this does not fix
the problem, return the monitor to Nellcor.
One or more display segments does not work.
• A User Interface PCB component has
failed.
Inspect the User Interface PCB for cracking, burning, or
damage. If damage is found, replace the User Interface PCB.
See
SpO2 PCB or User Interface PCB Replacement
on
page 50.
• The User Interface PCB is defective.
If damage is found, replace the User Interface PCB. See
SpO2 PCB or User Interface PCB Replacement
on page 50.
Pulse rate and %SpO
2
value does not display when the monitor is turned on.
• The SpO
2
sensor is not connected
properly.
Reconnect the SpO
2
sensor to the monitor.
Connect the SpO
2
sensor to the patient. See the sensor
Directions for Use.
• The SpO
2
sensor has failed.
Replace the SpO
2
sensor.
• The User Interface PCB is defective.
Replace the User Interface PCB. See
SpO2 PCB or User
Interface PCB Replacement
on page 50.
• The SpO
2
PCB is defective.
Replace the SpO
2
PCB. See
SpO2 PCB or User Interface
PCB Replacement
on page 50.
The display backlight does not come on when you press the Backlight button.
• The flex circuit between the front panel
and the SpO
2
PCB is damaged, or not
making a good connection.
Caution: Unlock J4 before attempting to
remove and reinsert the flex cable.
Inspect the flex cable for damage. Inspect the connection to
J4. If there is damage to the flex cable, replace the front
case. Otherwise, remove the cable from J4 and reconnect it.
Reassemble the monitor and retest. See
Separating the Front
and Back Cases
on page 48 and
Front Case Assembly
Replacement
on page 54.
Table 4:
Troubleshooting Guide
(Continued)
Condition
Recommended Action
Содержание OxiMax N-65
Страница 1: ...N 65 ...
Страница 6: ...iv OXIMAX N 65 Service Manual Contents ...
Страница 44: ...Performance Verification 38 OXIMAX N 65 Service Manual ...
Страница 66: ...Disassembly Guide 60 OXIMAX N 65 Service Manual ...
Страница 70: ...Spare Parts 64 OXIMAX N 65 Service Manual ...
Страница 74: ...Packing for Shipment 68 OXIMAX N 65 Service Manual ...
Страница 86: ...Specifications 80 OXIMAX N 65 Service Manual ...
Страница 94: ...Blank Page ...
Страница 95: ...Technical Discussion OXIMAX N 65 Service Manual 89 Figure 27 SpO2 PCB Schematic Diagram Sheet 1 of 7 ...
Страница 96: ...Blank Page ...
Страница 97: ...Technical Discussion OXIMAX N 65 Service Manual 91 Figure 28 SpO2 PCB Schematic Diagram Sheet 2 of 7 ...
Страница 98: ...Blank Page ...
Страница 99: ...Technical Discussion OXIMAX N 65 Service Manual 93 Figure 29 SpO2 PCB Schematic Diagram Sheet 3 of 7 ...
Страница 100: ...Blank Page ...
Страница 101: ...Technical Discussion OXIMAX N 65 Service Manual 95 Figure 30 SpO2 PCB Schematic Diagram Sheet 4 of 7 ...
Страница 102: ...Blank Page ...
Страница 103: ...Technical Discussion OXIMAX N 65 Service Manual 97 Figure 31 SpO2 PCB Schematic Diagram Sheet 5 of 7 ...
Страница 104: ...Blank Page ...
Страница 105: ...Technical Discussion OXIMAX N 65 Service Manual 99 Figure 32 SpO2 PCB Schematic Diagram Sheet 6 of 7 ...
Страница 106: ...Blank Page ...
Страница 107: ...Technical Discussion OXIMAX N 65 Service Manual 101 Figure 33 SpO2 PCB Schematic Diagram Sheet 7 of 7 ...
Страница 108: ...Blank Page ...
Страница 109: ...Technical Discussion OXIMAX N 65 Service Manual 103 Figure 34 User Interface PCB Schematic Diagram Sheet 1 of 3 ...
Страница 110: ...Blank Page ...
Страница 111: ...Technical Discussion OXIMAX N 65 Service Manual 105 Figure 35 User Interface PCB Schematic Diagram Sheet 2 of 3 ...
Страница 112: ...Blank Page ...
Страница 113: ...Technical Discussion OXIMAX N 65 Service Manual 107 Figure 36 User Interface PCB Schematic Diagram Sheet 3 of 3 ...
Страница 114: ...Blank Page ...
Страница 117: ...Index OXIMAX N 65 Service Manual 111 Y year 13 ...
Страница 118: ...Index 112 OXIMAX N 65 Service Manual ...
Страница 119: ......