Performance Verification
24
6. The N-550 will display:
•
75 %SpO2 (pass criteria is 73 to 77 %SpO
2
)
•
60 bpm
•
alarm
•
pulse amplitude indicator - low level modulation
Test #3: Modulation Level
1. Press the SRC-MAX % Modulation selection button. The SRC-MAX %
Modulation
LED will light.
2. The N-550 pulse amplitude indicator will fill (10 bars) with each pulse
beat.
3. The N-550 will display:
•
75 %SpO2 (test pass criteria is 73 to 77 %SpO
2
)
•
60 bpm (test pass criteria is 57 to 63 bpm)
•
alarm
•
pulse amplitude indicator - high level modulation
on page 22. The pulse amplitude indicator should
indicate high level modulation.
on page 23. The pulse amplitude indicator should
indicate high level modulation.