4-7
Section 4: Service Mode
4.3.1
Default Alarm Limits
The service technician can change the default alarm limits for either
patient mode (Adult or Neonatal) as described in Table 4-3: "Changing
Default Alarm Limits".
The monitor has the following alarms with adjustable level settings:
• EtCO
2
high and low levels (mmHg)
• FiCO
2
high level (mmHg)
• SpO
2
high and low levels (%)
• Pulse rate high and low levels (bpm)
Warning: Changing default settings will permanently change the
user’s start-up value and should be verified with the responsible
clinical personnel.
Table 4-3: Changing Default Alarm Limits
Objective
Action
Response
To access the Default
Alarm Limits menu
after entering last
password digit
short press
To change the patient
mode
short press
To access a new
parameter
short press
To change the value
short press