Do not use in oxygen rich environments, in the presence of propane, 
gasoline or other combustible gasses, or near dust and debris. Doing so may 
cause explosion. 

 

Do not leave the illumination of for prolonged periods or cover the device 
during illumination. Doing so may cause the device to reach high 
temperatures. 

 

Never touch liquid which may have leaked from the Battery Pack. In case of 
contact with the eye(s), flush with clean water without rubbing, and consult a 
physician immediately. If leaked liquid gets on skin or clothing, wash with 
clean water immediately. Failure to do so may result in serious injury, 
including loss of eyesight. 

 

Do not short-circuit connectors or touch them with bare hands. Doing so may 
cause electric shock or device failure. 

 

Do not attempt to disassemble or modify the device. Doing so may result in 
electric shock, fire or device failure. 

 

Never remove the cover of the battery pack. Attempting to charge 
commercial batteries so may cause leakage of liquids, heat generation or 
rupturing during charging. 

 

Do not touch the device with wet hands. Doing so may result in electric 
shock. 

 

In the event of ingress of water or other liquids, unplug from the power outlet 
immediately. Failure to do so may result in electric shock, fire or device 
failure. 

 

Unplug the power cable from the power outlet prior to cleaning. Failure to do 
so may result in electric shock 

 

Take care to only touch insulated parts when inserting the power plug into 
the power outlet. Failure to do so may cause electric shock or fire. 

 

 

 

Do not sterilize the device. Doing so may cause device deformation or 
failure. 

 

Do not install the device in places where it may be difficult to plug in/out 
to/from the power outlet or to switch power since the plug-out from the outlet 
is one of the separation methods between the device and the power supply 

 

Do not use in excessively humid or saline environments, or places where the 
device may be sprayed with water. Doing so may result in electric shock or 
device failure. 

 

Do not expose the device to direct sunlight or harmful radiation. Doing so 
may lead to unexpected heating and result in device failure. 

 

Do not attempt to charge the device in temperatures exceeding 35 °C 
(95 °F). Doing so may prevent the device from charging or result in device 
failure. 

 

Do not attempt to charge the device when wet. Doing so may cause 
excessive heating, fire or rupture. 

 

Do not place heavy objects on, apply excessive tension to, or fold cables, 
cords and connectors. Doing so may result in damage and electric shock or 
fire. 

 

Do not use paint thinners, cleaning agents or boiling water to clean the 
device. Doing so may cause deformation or device failure. 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Instruction

 

Instruction

 

Instruction

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Содержание IO-A LED

Страница 1: ...is is a battery powered device to examine the inside and fundus of the eyes by illuminating with an LED bulb The device is powered by the Battery Pack IO BP which can be attached to the Headband or a...

Страница 2: ...ng the IO BP to the Scope Unit 11 3 2 Operation 12 3 2 1 Wearing the IO 12 3 2 2 Positioning the Scope Unit 12 3 2 3 PD Adjustment 12 3 2 4 Illumination 12 3 2 5 Examination 13 3 2 6 Using the Teachin...

Страница 3: ...and caution you of potential hazards to yourself and others Potentially hazardous situation which if not avoided could result in death or serious injury Potentially hazardous situation which if not av...

Страница 4: ...ses Catalog number product code Serial number 1 3 Safety Information For your safety please comply with the following precautions Do not use battery packs or AC adaptors other than the standard access...

Страница 5: ...ing Failure to do so may result in electric shock Take care to only touch insulated parts when inserting the power plug into the power outlet Failure to do so may cause electric shock or fire Do not s...

Страница 6: ...ned off during transport or storage and only use the specified casing Failure to do so may result in deformation device failure or fire When disconnecting cables be sure to hold the plug Never pull th...

Страница 7: ...do not use the IO and contact your local dealer immediately Set Type Item Set 4 Set 3 Set 2 IO Main Unit See 2 2 1 1 1 1 IO BP See 2 2 2 1 1 1 AC Adapter example 1 1 1 Extension Cord 1 m 1 1 1 Belt Ho...

Страница 8: ...nation field 7 Aperture Selection Marker 8 Eyepiece A 2 D diopters lens is included 9 PD Adjuster Adjusts the pupil distance PD The scale 54 mm to 74 mm is indicated above the Eyepiece 10 Illumination...

Страница 9: ...ds the Scope Unit Can be flipped up 17 Overband Fixation Knob Loosen to flip up the Overband Tighten to keep the Overband in place 18 Headband Cover s Covers the adjustment mechanics of the Headband F...

Страница 10: ...ff and allows for continuous intensity adjustment 4 Illumination Off Mark 5 Illumination Intensity Mark 6 Battery Lamp Indicates the remaining battery capacity 7 Indicates power output from the Lamp J...

Страница 11: ...ange 3 Charging time is approximately 2 5 hours from a fully discharged state Note The IO cannot be illuminated while the DC Plug is connected 4 Once fully charged the CHARGE Lamp will light green 5 U...

Страница 12: ...th a click see Fig 3 4 Note The plug and jack interlock at a single configuration Rotate the plug as needed to connect smoothly 2 When using the Belt Hook the Lamp Cord attached to the Scope Unit is n...

Страница 13: ...n the Scope Unit Fixation Knob to fix in the position with the best view when observed through the eyepieces see Fig 3 7 3 2 3 PD Adjustment 1 Turn the Observation Angle Adjustment Lever upward to ful...

Страница 14: ...lens in one hand direct the illumination into the patient eye It is recommended to place the unoccupied fingers of the hand holding the aspherical lens on the patient s face for stability 2 Bring the...

Страница 15: ...observation axes remain apart see Fig 3 9c Examining Vertically Elliptical Pupils When examining the fundus through vertically elliptical pupils raise the Illumination Angle Adjustment Lever and lowe...

Страница 16: ...detaching the Teaching Mirror from the Scope Unit depress the mark on both the upper and lower faces of the Teaching Mirror and pull gently Note Do not attempt to detach the Teaching Mirror forcefull...

Страница 17: ...pe the surface using a dry lint free cloth or cloth moistened with diluted neutral detergent and then wipe dry Use a cloth moistened with alcohol to disinfect 4 2 Replacing the LED Bulb As LED bulbs h...

Страница 18: ...operate If such events should occur contact your local dealer to purchase a new Battery Pack 4 4 Disposal Dispose of the IO and its accessories excluding the IO BP in accordance with your local regula...

Страница 19: ...oned Move the lever s until you feel a click Observation image blurry Foreign matter is adhered to the eyepiece or observation system Clean in accordance with 4 1 1 Foreign matter is adhered to the as...

Страница 20: ...tion Degree of protection against electric shock Internally powered ME equipment Applied parts No applied parts Degree of protection against harmful ingress of water or particulate matter IPX0 Method...

Страница 21: ...ort contact your local dealer or Neitz located at the following address LOCAL DEALER Neitz Instruments Co Ltd 4F Ichibancho Court 15 21 Ichibancho Chiyoda ku Tokyo 102 0082 Japan Phone 81 3 3237 0552...

Страница 22: ...Ophthalmic instruments Fundamental requirements and test methods Part 2 Light hazard protection indicates that the IO is classified to be Group 2 and ophthalmic instrument for which a potential light...

Страница 23: ...O is intended for use in the electromagnetic environment specified below The customer or the user of the IO should assure that it is used in such an environment Emissions Test Compliance Electromagnet...

Страница 24: ...000 4 5 1 kV line s to line s 1 kV line s to line s Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power...

Страница 25: ...d RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with t...

Страница 26: ...cy of transmitter m 150 kHz to 80 MHz 1 2 80 MHz to 800 MHz 1 2 800 MHz to 2 5 GHz 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maxi...

Страница 27: ...25 25 Notes...

Страница 28: ...Japan Phone 81 3 3237 0552 FAX 81 3 3237 0554 E mail int neitz co jp URL http www neitz ophthalmic com Distributed by As part of our policy of continued product improvement we reserve the right to alt...