49
NED
XCM4085DLMT4
UME-0058-03
4.8.3
External Trigger Exposure Mode (Trigger Level)
In external trigger exposure mode (Trigger Level), the exposure time is determined
by the high trigger pulse time, each exposure starts with the rising edge and the line
period is determined by the time rising edge to rising edge of trigger pulse. The range
of programmable exposure time, the timing chart of the exposure and the readout are
shown below.
Table 4-8-3-1 Programmable Exposure Time (Trigger Level)
r
Readout time
4096pixels_4tap
12.05
4096pixels_2tap
24.09
2048pixels_2tap
12.05
4096pixels_4tap
(
TypeB
)
12.05
a
Trigger pulse High time
4096pixels_4tap
≧
11.072
4096pixels_2tap
≧
23.882
2048pixels_2tap
≧
11.382
4096pixels_4tap
(
TypeB
)
≧
11.072
b
Trigger pulse Low time
≧
2.048
c
Trigger pulse cycle
4096pixels_4tap
≧
13.12
4096pixels_2tap
≧
25.93
2048pixels_2tap
≧
13.43
4096pixels_4tap
(
TypeB
)
≧
13.12
(unit
:
µs)
Figure 4-8-3-1 External Trigger (Trigger Level) Exposure Mode
at 4096pixels_4tap
Note:
The timing of reading out does one scanning delay from the exposure.