English-16
4. Showing the test results
When the measurement is completed, the test results are shown.
Test results
A. Test
results
Passed: Max. deviation
≤
10%
Failed: Max. deviation > 10%
Canceled: Test has been canceled.
Passed(!),
Failed(!)
“(!)” following Passed and Failed indicates that the recommended condition has not been met. (The test was
performed in less than 60 minutes of continuous lighting time of the backlight.)
If the test is continuously performed in less than 30 seconds, “Unstable” is shown in LUMINANCE and “(!)” is
shown in RESULT.
B. Maximum
deviation
The test results are saved until the next test or calibration.
If the achieved maximum deviation does not conform to the standard even though the test has been performed under the
recommended condition, calibrate the display using an external calibration sensor.
Without the OSD operation for over 15 seconds, the OSD disappears automatically.
Advanced OSD Controls without video input
If no video signal is available, the display enters the power save mode and the OSD cannot be displayed.
In this case, please follow the procedure below.
1. Confi rm that the display is in power save mode and the LED next to the power switch is lit orange. If the LED is off, please
push the power switch to turn on the monitor.
2. Hold down the (+) plus button and the (-) minus button simultaneously for over 2 seconds.
The “INFORMATION” OSD menu opens on the screen. The LED color changes from Orange to Green.
3. Switch EDID, input source, Confi guration or DICOM conformance test referring to this manual.
How to Change the EDID settings (see page 12)
How to Change the input source (see page 13)
How to Change Confi guration (see page 14)
How to make DICOM conformance test (see page 15)
4. If the display enters the power save mode during the change process, please start from the beginning again.
NOTE:
This mode is released after any of the following conditions are met:
• 3 hours has passed.
• The display power has been turned off.
• The DICOM conformance test has been performed.
Содержание MD211G5
Страница 1: ...MD211G5...
Страница 3: ...English...