EasyOne Pro™ / EasyOne Pro™ LAB
105
Hygiene, Cleaning, Maintenance, Disposal
Three full trials are required for the
Single Flow
calibration check,
whereas three tests comprising three trials each with different flow rates
are required for the
Multiflow
calibration check.
After you perform the maneuver, you will see the text
Accuracy confirmed
and, beneath it, the percentage deviation and the average flow velocity of
the pump stroke.
You can repeat the test, print the result or quit the program. The calibration
test remains stored and can also be viewed or printed out later.
If you do not reach ±3% accuracy, please follow the troubleshooting instruc-
tions in chapter
on page 130. Should you not be
able to remedy the defect by following these instructions either, please con-
sult
ndd
or one of their representatives.
DLCO Calibration Check
The American Thoracic Society (ATS) recommends checking DLCO calibra-
tion on a weekly basis or whenever irregularities are suspected. DLCO cali-
bration can be checked with a DLCO simulator (see following section) or
with a test subject (see
"Calibration Check with Test Subject"
In addition, the ATS/ERS guidelines (Literature [11], page 182) describe a
third method for checking DLCO calibration for which a 3-L syringe is used.
Fig. 7-3
Calibration syringe