Troubleshooting Tips
130
EasyOne Pro™ / EasyOne Pro™ LAB
9
Troubleshooting Tips
Should you encounter problems operating your system, please consult the
table below for troubleshooting tips.
Note
The most recent information and frequently asked questions
can be found on our web site at www.ndd.ch.
When the
EasyOne
Pro™
is switched
on, you see the fol-
lowing error mes-
sage on the display
Self-test failed
Quit the program and
restart. If you receive
the same message
again, contact your
ndd dealer.
When you start a
test, you see the
message
Check
spirette insertion.
The
spirette™
is not
correctly positioned.
Ensure that the trian-
gle on the spirometer
is lined up with the tri-
angle on the
spirette™
Calibration check
outside ±3.5%.
The
spirette™
is not
correctly positioned.
Insert the
spirette™
as
described in section
"Inserting the One-Way
Valve, DLCO barri-
ette™/FRC barriette™
and spirette™"
You have not used an
ndd adapter.
Use the ndd calibra-
tion adapter.
There are leaks in the
syringe connection.
Check the connec-
tions.
The specified syringe
volume differs from
the actual syringe vol-
ume.
Choose the correct
syringe volume via
Configuration
>
Test
>
Cal-Check
.