Pro Advantage® 

Urine Analyzer 

 

 

 

Operator’s Manual

 

 
 
 
 
 
 

Содержание Pro Advantage

Страница 1: ... Pro Advantage by NDC All rights reserved PS 733 Rev 00 3 10 1 23 Pro Advantage Urine Analyzer Operator s Manual ...

Страница 2: ...printing However the manufacturer continues to update and change this manual The manufacturer reserves the right to change the content of this manual at any time without notice If the manner in which this machine is used differs from the procedures specified in this manual then the instrument and the results may be impaired ...

Страница 3: ... to change strip type 11 4 Analyzer Set Up 12 How to change your urine analyzer settings 12 Printer Settings 12 Urine Color Settings 12 Language Settings 13 Baud Rate 13 How to feed printer paper through the printer head 13 How to change LCD screen contrast 13 5 Troubleshooting 14 Error Messages 14 Errors which Disable the Instrument 14 Errors which do not Disable the Instrument 14 Calling for Ass...

Страница 4: ...33 Rev 00 3 10 4 23 Appendix 2 Default Settings 20 Appendix 3 Analyzer Summary Policy 21 Intended Use of Analyzer 21 Optical System Explanation 21 Internal Checks Explanation 21 Human Eye vs Instrumental Optics 21 Appendix 4 Warranty 22 10 Index 23 ...

Страница 5: ...e it is turned on How does the analyzer work The user has the option to input the Patient ID and also select the urine color before running the patient sample Testing starts when Urine Test is selected on the main menu screen From this point you have 5 seconds to dip the urine reagent strip blot the edge of the strip and place it on the test strip bed The analyzer will then count down from 60 seco...

Страница 6: ...icates the symbols that appear on the exterior packaging of the Pro Advantage Urine Analyzer the analyzer itself the power adapter and the urine reagent strips Indicates a power on off button In vitro diagnostic device Temperature Requirements for Operation 15º 30º C Expiration Date MM YYYY Serial Number Lot Number ...

Страница 7: ...32 connection located on the back of the analyzer This requires a 9 pin serial cable that can be purchased separately at an electronics store or from your Local Technical Support Providers and Distributors Loading the Printer Paper or Label Roll 1 Remove the printer cover by pulling up on the tab 2 Open the thermal paper roll compartment cover by pressing down on its tab and pulling away from the ...

Страница 8: ...ing results Strip Type This menu option allows for the user to change the strip type that will be used on Figure 1 Front View of Pro Advantage Urine Analyzer the analyzer Color Selection This menu option allows for the user to select the color of the patient s urine color prior to testing Data Output Serial Port The output of data uses a serial port which requires a 9 pin serial cable for transmis...

Страница 9: ...g period the analyzer will move the strip bed into the analyzer to begin reading the color of each parameter Once the testing time has expired the results will appear on the screen see Fig 6 The analyzer will also print the results automatically if the analyzer has been set up to automatically print results The date time patient ID and color will be printed along with the test results see Fig 7 Th...

Страница 10: ...digit patient ID see Fig 9 3 Use the selection keys to choose a numeric digit from 0 9 4 Press the YES key to highlight the second digit of the Patient ID 5 Continue to repeat steps 3 and 4 to finish entering the 6 digit Patient ID Once the Patient ID is selected you may choose to enter the urine color by highlighting Color Selection see How to enter urine color on this page If you choose not to s...

Страница 11: ... s in its internal memory The Pro Advantage Urine Analyzer also allows the user access to these stored Patient ID s To access the past Patient ID results the user must start at the Main Menu 1 Use the selection keys to highlight the Patient ID option see Fig 8 2 After highlighting the Patient ID press the YES key to enter the patient ID which you need to recall see how to enter a Patient ID on pag...

Страница 12: ...19 depending on what function the analyzer was previously performing PrintOFF The analyzer will not print via the on board thermal printer PrintON The analyzer will print via the on board thermal printer If you choose to connect a dot matrix style external printer and have the print function turned off this analyzer will print via the external printer Figure 19 Print Option Turned Off Figure 20 Ur...

Страница 13: ... selection keys to select the Baud Rate required see Fig 24 4 Press NO to return to the previous screen Figure 24 Baud Rate Settings Screen Figure 25 Paper Feed Option How to feed printer paper through the printer head The Pro Advantage Urine Analyzer requires a thermal paper roll to be used to properly print the results of each patient sample At times it may be difficult to feed the paper through...

Страница 14: ...nalyzer Calling for Assistance If your Pro Advantage Urine Analyzer is displaying an error please follow the instruction shown in this manual to resolve the error shown on the analyzer If the error still exists after following the steps in the manual please contact your local technical support provider If you are calling for assistance to a displayed error please have the following items ready 1 D...

Страница 15: ...s a CLIA waived analyzer when used with the manufacturer s branded urine reagent strips For CLIA waived settings we advocate for the use of positive and negative controls with each new bottle opened Water should NOT be used as a negative control Contact your manufacturer representative for any supplementary information about quality control testing ...

Страница 16: ...er 3 Remove the strip bed from the analyzer 4 Insert strip bed into the solution making sure the strip bed is submerged 5 Submerge the strip bed for a minimum of 2 minutes and maximum of 5 minutes 6 Do not submerge longer than 5 minutes 7 Rinse the strip bed thoroughly with water 8 Use a dry soft cloth and replace strip bed into analyzer as previously described on this page in Routine Cleaning of ...

Страница 17: ...e Analyzer has the ability to change the units in which the results are displayed and printed 1 Start at the Main Menu screen and hold the DOWN selection key and press the NO key 2 The units will change from C1 to C2 to S1 or S2 see Fig 24 and 25 C1 English Conventional Normal System C2 English Conventional Plus System Figure 24 Conventional Units Standard Figure 25 Conventional Units Plus ...

Страница 18: ... has the ability to change the units in which the results are displayed and printed 1 Start at the Main Menu screen and hold the DOWN selection key and press the NO key 2 The units will change from C1 to C2 to S1 to S2 see Fig 26 and 27 S1 English International Normal System S2 English International Plus System Figure 26 Conventional Units Standard Figure 27 Conventional Units Plus ...

Страница 19: ...Screen up to 45 degree incline Weight Pro Advantage Urine Analyzer without power supply 2 36 kg 5 21 lb Operating Temperature 18 C to 30 C 64 F to 86 F Operating Humidity 18 to 80 Relative Humidity Optimal Operating Temperature 22 C to 26 C 72 F to 79 F Optimal Operating Humidity 35 to 55 Relative Humidity Optimal ranges ensure that the reagent results are optimized for performance and stability A...

Страница 20: ...lude Color Selection in Results Enabled Enabled Date Format Language MM DD YYYY English Time Format Results Format Units Selection 12 hour Conventional Plus System Mark Positive Results Disabled Enabled System Settings Connectivity Allowing Results to be Sent to Computer Printer Automatic Enabled Contrast Setting Baud Rate 0 zero 115200 Parity Stop Bits None 1 one Strip Type 10 Parameter ...

Страница 21: ...cted radiation of the strip block to testing wavelength Ts The intensity of reflected radiation of the strip block to reference wavelength Cm The intensity of reflected radiation of the reference block to testing wavelength Cs The intensity of reflected radiation of the reference block to reference wavelength Internal Checks Explanation When the user first turns on the Pro Advantage Urine Analyzer...

Страница 22: ...airs or replace any parts that are necessitated by abuse accidents alteration misuse neglect maintenance by anyone other than the manufacturer or failure to operate the instrument in accordance with instructions Further the manufacturer assumes no liability from malfunction or damage to the manufacturer instruments caused by the use of reagents other than reagents manufactured or recommended by th...

Страница 23: ...age 7 Contrast LCD Page 11 Connecting to a PC Page 5 Q Quality Control Page 13 D Date and Time Page 6 and 10 R Default Settings Page 19 Results Page 7 Display Page 18 Dimensions Page 17 S Serial Port Page 6 E Specifications Page 17 Error Messages Page 12 Strip Type Page 6 and 9 System Settings Page 10 11 F Strip Bed Page 6 G T Testing Page 7 H Troubleshooting Page 12 I U Units Page 15 16 J Convent...