The Calibration (CAL) Mode menu allows scale
calibration. The items in the Calibration menu are
as follows:
Pounds/Kilograms (lb or 1000G)
– Selects
the unit of measure of your calibration test
weights.
Scale or Class
– When calibrating the scale
for lb, calibrate the unit as a SCALE or as a
CLASS (weight classifier/postal rounding).
U On / U Off
– When configured for U ON,
the scale will allow switching between lb/kg
using the
UNITS
key.
Capacity (100, etc.)
– Selects the capacity of
the scale.
Follow these steps to calibrate the scale. Refer to
Figure 2.
1. From normal weighing mode, move Switch 1
to the Menu Mode or OPEN position…
DIAG
is displayed. Press the
ZERO
key
until
CAL
is displayed.
2. Press the
UNITS
key to start calibration…
LBS
or
1000G
(kg) is displayed.
3. Press the
ZERO
key to toggle between the
choices of unit of measure (lb or kg). When
the choice you want is displayed, press the
UNITS
key to accept…
The choice is accepted and
U ON
is
displayed.
12
Calibration Mode
Step-by-Step
Instructions for
CAL Mode
Warning!
Entering this mode can
erase the calibration
already saved. Special
calibration weights are
needed to use
calibration mode.
Содержание 7821
Страница 1: ...User s Manual Model 7821 Bench Scale...
Страница 8: ...Model 7821 Bench Scale User s Manual 8 Figure 2 Menu Structure...
Страница 24: ...Model 7821 Bench Scale User s Manual 24 Notes...
Страница 25: ...Model 7821 Bench Scale User s Manual 25 Notes...
Страница 26: ...Model 7821 Bench Scale User s Manual 26...
Страница 27: ...Model 7821 Bench Scale User s Manual 27...