Model 7620U Bench Scale User’s Manual
18
Division size for a
scale calibrated
as a:
6-30:
0-6 kg x 0.001 kg
6-30 kg x 0.005 kg
5-30:
0-5 kg x 0.001 kg
5 -30 kg x 0.005 kg
6-35:
0-6 kg x 0.001 kg
6-35 kg x 0.01 kg
5-50:
0-5 kg x 0.005 kg
5-50 kg x 0.01 kg
If this procedure is
attempted without any
calibration weights
applied, the scale will
abort the process and
retain the original
calibration data.
Close Switch-1 if you
don’t have correct
calibration weights.
2. Press the
F1
key to start calibration…
The current capacity setting is displayed.
Use the
ZERO
key to scroll through the
choices of capacities:
“5-30, 6-30, 6-35,
5-50.”
3. When the correct capacity selection is
displayed,press the
F1
key…
The scale prompts
LOAD 0
.
4. Clear all weight from the scale platter and
press the
F1
key…
After a brief wait
“LOAD XX”
(XX=the
span weight) is displayed. Alternate
calibration points can be chosen using
the
ZERO
key to toggle between
choices.
5. Place the chosen (alternate) calibration
weight on the scale and allow it to stablize
and then press the
F1
key…
After a brief wait
“DONE”
is displayed.
The scale then prompts
“CAL.”
6. Remove the calibration weight and press the
ZERO
key…
“DIAG”
is displayed.
OR…
7. Return Switch 1 to the closed position…
The scale returns to normal weighing
mode.
The scale is now tested, configured and cali-
brated. It is ready for use in your application.