It is important that the Resolv4K Zoom be properly installed onto the mount. All flat mounts
(1-80370, 1-80371, 1-80270, 1-80271) have a ledge on the backplate. This ledge is to sit
within the notch located at the center of the Resolv4K Zoom or Fixed body tube (see diagram
below). This provides a solid surface for which the system can be held. Install the saddle
over the lens and secure with M5 screws using the 4mm hex wrench supplied with the mount.
USING THE MOUNT WITH THE RESOLV4K SYSTEM
Left:
Location of ledge on mount and notch on body tube.
Right:
Secure saddle with screws.
To set parfocality zoom to high magnification. Adjust
the “Z” axis to achieve best focus. Then zoom to
low magnification, do not adjust “Z” axis to refocus.
Unlock the screws and adjust the parfocality using the
adjustment ring on the rear adapter. When the image
is in optimal focus, you should be able to go through
zoom without loss of focus. Lock the parfocality ring
in two places (see Fig.1) using a 1.5mm hex wrench
(supplied with your lens).
All Resolv4K rear adapters have boresight
adjustability. The adjustment is made by using a
1.5mm hex wrench to adjust the M3 set screws at the
top of the adapter (see Fig. 2).
Tip: For easier boresight adjustment, first loosen the
M3 screws and align the “X” axis of the sensor along
the same “line” as one of the M3 screws. Next, use
the (2) wrenches (supplied with your lens) in opposite
positions along the “X” axis to adjust the boresight.
This type of alignment enables the user to “push” the
image in the direction being actuated.
SET PARFOCALITY & BORESIGHT
Fig. 1: Location of parfocality
adjustment. Secure with set screw.
Fig. 2: Location of set screws for
boresight adjustment.
Содержание Resolv4k Series
Страница 1: ...Installation Instructions...