background image

NA

Y

AC 

9. Troubleshooting

Fault 

Possible reason 

1.  Wiring is malfunctioning. 

2.  ogeration voltage is

a  normal. 

3.  Motor is malfunctioning. 

4. Overload protector is active. 

Pump cannot 

5. Oil  temperature  is  below 10

°

start 

6. Pump is jammed. 

7. Out of operating for long. 
8. Oil is too viscous. 

9. Exhaust  filter or exhaust 

line is clogged. 
10. Pump inner parts are 

damaged. 
1. Vacuum system 

configuration is unreasonable. 

2. Vacuum system leak. 

3.  Measuring technique or

gauge is unsuitable. 
4.  Poor lubrication. 

Pump cannot 

4.1 Oil filter is obstructed. 

reach ultimate  4.2 Oil is unsuitable. 

pressure. 

4.3 Oil channel is obstructed. 

4.4 Oil is insufficient. 

5. Anti-suckback oil valve is 

malfunctioning. 

6. Intake line is dirty. 

1. Intake port line is clogged. 

2.  Connecting lines are too 

Pumping 

narrow or too long. 

speed is too 

3.  Exhaust line is clogged.

low. 

4. Oil mist filter is clogged.

5. Anti-suckback oil valve is

malfunctioning.

1 .  Operation voltage is 

abnormal. 

2.  Foreign matters entering

Noise is 

into the pump. 

abnormal. 

3. The oil level of the pump 

is too low. 

4. The pump inner parts are 

damaged.

1. Poor ventilation.

2. The fan is damaged.

3. Pumped gas temperature

is too high.

Pump gets 

4.  Poor lubrication.

hotter than 

4.1 The  oil filter or exhaust 

usually 

line is clogged. 

observed. 

4.2 Oil is unsuitable. 

4.3 Oil channel is obstructed. 
4.4 Oil is insufficient. 

5. The condensator is dirty. 

6. Ambient temperature is

too high.

10 

Remedy 

1.  Check and  repair wiring. 

2. Make sure the voltage is within rated voltage's tolerance of ±10%. 

3.  Replace the motor. 

4. Check the ambient temperature or pumoed gases temperature.

5. Increase the ambient temperature. 

6. Repair the pump. 

7.  Repair the pump. 

8. Change the oil. 
9. Replace the filter or clean the exhaust line. 

10. Repair the pump. 

1. Choose a suitable pump. 

2.  Check the system. 

3. Use correct measuring technique and  gauge. Measure the 

pressure directly at pump's intake port. 

4.1 Change the oil filter. 

4.2 Change the oil. 

4.3 Clean the channel. 

4.4 Add the oil to the level as requested. 

5.  Repair anti-suckback oil valve. 
6. Clean vacuum lines. 

1. Clean the  intake port line.
2.  Use adequately wide and short connecting lines.

3.  Keep exhaust port line unobstructed.

4. Change the oil mist filter.

5.  Repair anti-suckback oil valve.

1.1 Check the power supply,  switches and wiring connection. 

1 .2 Make sure the voltage is within rated voltage's tolerance of ±10%. 
2. Repair the pump. 

3. Add the oil to the level as requested.

4. Disassembly and replace the parts.
1. Improve ventilation environment.

2. Change the fan.

3. Add cold trap at intake port.

4.1 Replace oil filter or clean the exhaust line. 
4.2 Change the oil. 
4.3 Clean the channel. 

4.4 Add the oil to the level as requested. 

5. Clean the condensator.
6. Reduce ambient temperature.

Fault 

Possible reason 

Remedy 

1. Oil  comes from the

1. Check the vacuum system.

vacuum system.

Oil in the 

2. Anti-suckback valve

intake line or in  spring is obstructed. 

2.  Replace anti-suckback valve spring.

vacuum vessel.  3. Anti-suckback valve plate

is obstructed. 

3. Replace anti-suckback valve plate.

Vacuum pressure  4.  Oil level is high.

4. Drain the excess oil.

in system rises 

1.  Vacuum system leak.

1. Check the system.

too fast when 

2. Anti-suckback valve is

pump stops. 

malfunctioning.

2.  Repair anti-suckback valve.

Too much oil 

1.  Too much oil in  the pump. 

1. Drain some oil.

in the exhaust 

2.  Continuous operation under 

2. Shorten exhaust time.

port. 

high pressure in the intake port. 

3. Oil mist filter is obstructed. 

3.  Replace oil  mist filter.

Sealing 

1. The seal  is damaged.

1. Replace the seal.

surface leak 

2. Seal ring is damaged.

2. Replace the seal ring.

Table4 

10. Warranty

NRD series vacuum pumps have one year guarantee from the date  of  purchase.  Spare parts will be provided free of charge within 

the period of guarantee in condition that the pump is operated according to the operating manual. The following failures will be charged 
for repair: 

1)  Malfunction caused by nature disasters or artificial factors 

2) Malfunction caused by special operating environment 
3) Damage of wear parts
4) Malfunction caused by abnormal operation or incorrect use indentified by our engineers

Before the pump is delivered to our company for maintenance, it is required to acquire the information if the pump is contaminated
(and which kind of pollutants) or hazardous substances exist inside the pump. Without receiving this statement in advance, the pump

will be returned to the address of the sender. 

11. Accessories

To  ensure the stability of the pump, it is recommended to use genuine spare  parts and accessories provided by the manufacturer. 

If you need further assistance,  please contact us. 

The accessories below are for your selection: 

1) Wear parts. Check the exploded view and spare parts list to select the required spare parts with corresponding item numbers. 

2) Intake port/exhaust port other than default one 
3) Dust filter 

11 

Содержание NRD Series

Страница 1: ...C NRD Series User Manual Dual Stage Oil sealed Rotary Vane Vacuum Pump NRD16 16T l E _ 5 Commonwealth Ave Woburn MA 01801 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot...

Страница 2: ...ury Notice Failure to observe the terms could result in damage to the pump This warning label indicates risk of electrical shock Disconnect the pump from the power supply before beginning with connect...

Страница 3: ...tive gas within the scope of low vacuum and a small amount of non condensable gas The pump is specially suitable for coating refrigeration freeze drying sterilization instruments analysis leak detecti...

Страница 4: ...D4 Without gas ballast With gas ballast Pressure Pa Fig 3 5 Transport and Storage Warning Do not move the pump unless it s switched off and the power supply is disconnected Warning Make sure that the...

Страница 5: ...ich can shut off the entire product In the event that the thermal limiter operates nothing is need to do but waiting for the motor to restart automatically if it has been cooled to the temperature whi...

Страница 6: ...n the first time and follow all appropriate safety procedures If the potential hazard still exists the pump must be decontaminated before any maintenance 8 1 Routine inspection Content Operation Inter...

Страница 7: ...ithin rated voltage s tolerance of 10 2 Repair the pump 3 Add the oil to the level as requested 4 Disassembly and replace the parts 1 Improve ventilation environment 2 Change the fan 3 Add cold trap a...

Страница 8: ...01 Gas ballast 15 0 ring FKM 300310079 Support 16 Gasket Paper 320210101 Front chanter 17 Seal FKM 300280602 Front chanter 18 Spring SUS 320110104 Front rear rotor 5 NRD16 320100101 Front rotor 2 19 V...

Отзывы: