Important Recommandations :
In order to get a maximum efficiency of these sort of product,
it is necessary :
- to choose the appropriate size for the patient
- to get the best possible adjustment to the patient
These products must not be in direct contact with a wounded skin.
v 05.12/2020
THIS MEDICAL DEVICE MUST BE SET UP BY A PROFESSIONAL
OR THE USER MUST BE EDUCATED TO SET IT UP.
STANDARD SLING
Composition
100 % polyester fabric
40 mm polyester strap - 1450 DaN
50 kg/m3 PE foam
20/25 kg/m3 PE foam
20 mm polyester bias
CAUTION
Read the instruction manual before using the slings.
The backrest hangers must be hung on the same color.
Never change the attachment setting of the ears to the stand-up lift hooks,
when someone is installed in the sling.
Before lifting a person with the stand-up ligt, make sure the sling is securely
hooked to the hooks.
INSTRUCTIONS FOR USE
Do not use this sling if the size is not suitable for the patient.
Position the bottom of the backrest around the shoulder blades in order that
the ears of the attachment system are on either side of the person.
Place the fasteners of the backrest under the arms of the person, so that the
beads are under the armpits.
Close the plastron.
Hook the two fasteners of the backrest to the stand-up lift hooks.
SETTINGS
The fastening system, with multiple hangers, allows to adjust the position of
the person.
SIZE & DIMENSIONS (in cm)
B
A
Plastron
A
B
S
70 / 90
21
78
M
90 / 115
25
92
L
115 / 140
29
106
XL 140 / 165 33
120