Trex™ HD Monitoring Amplifier
User & Service Manual
12
Table 3 - Electromagnetic Immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The
Trex HD Monitoring
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Trex HD Monitoring
should assure that it is used in such an environment.
Immunity test
Test Level
IEC60601-1-2
Compliance
Level
Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Trex HD
Monitoring, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d=1.2 ×√P
150kHz to 80MHz
d=1.2 ×√P
80MHz to 800MHz
d=2.3 ×√P
800MHz to 2.5GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic Site survey
,
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
3
Field strengths from fixed transmitters such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the
Trex HD Monitoring
is used exceeds the applicable RF compliance level above, the
Trex HD
Monitoring
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
Trex HD Monitoring
.
4
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.