6-12
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Repair Procedures
Figure 6.4
Accessing the blower assembly
5
Lift the blower assembly and position it to allow access to the ground wire
screw on the motor side of the blower housing.
6
Remove and save the ground screw and star lock washer.
7
Disconnect the blower assembly at the inline quick disconnect terminals, then
remove the wire ties and old blower.
8
To replace only the motor, follow steps:
a
Remove and save the three screws that secure the inlet cover to the side of
the blower assembly (see Figure 6.5).
Figure 6.5
Blower assembly, exploded view
b
Remove the cover.
c
Within the blower assembly, locate the set screw that attaches the squirrel
cage to the shaft of the blower motor.
HEPA filter
FRONT
Acorn nuts and washers
Blower
Blower bracket
Blower filter plate
Wire cover screws
(2 on new units; 3 on
replacement motors)
Grounding wire
Blower housing
Inlet cover
Squirrel cage
Set screw
Molded rubber
mounting washer
Studs (x4 on new units;
x2 on replacement motors)
Wire cover