10
Troubleshooting
10.1
Powering
Problem
Cause
Solution
The On/Off light indicator is
not lit
•
There is no power supply
to the device.
•
Check that the cables of the power supply are
firmly connected to the device and the mains
outlet socket.
•
Check that the mains supply is switched on.
10.2
Software/device communication
Problem
Cause
Solution
During installation, the install-
ation process may be inter-
rupted. Various error messages
may occur.
•
The PC’s virus scan is active
and prevents the install-
ation process from pro-
gressing.
•
Deactivate the PC virus scan until the install-
ation process is completed.
When Otosuite is launched,
one or more error messages
may appear regarding XML
errors.
•
An earlier version of Oto-
suite was installed. An
extended version of data
sets has been installed
with the new version of
Otosuite.
•
Accept the error message. This message only
appears the first time the new version of Oto-
suite is launched.
There is no connection to the
device.
•
The USB cable connecting
the device to the PC was
connected prior to
installing Otosuite. This res-
ults in a Windows default
driver being allocated.
1. Select the Windows Device Manager followed
by Universal Serial Bus controllers.
The faulty connection will be marked by a yel-
low question mark in the list.
2. Uninstall the driver.
3. Make sure Otosuite is installed and relaunch
Otosuite.
Connection to the audiometer
was lost.
•
A problem occurred in the
USB cable communication.
1. Try to disconnect and reconnect the USB
cable.
2. If this does not work, power off and then
power on the device.
3. Contact technical support if the problem per-
sists.
84
Madsen A450
10 Troubleshooting