Warning
•
Indicates that there is a risk of death or serious injury to the user or patient.
Caution
•
Indicates that there is a risk of injury to the user or patient or risk of damage to data or the device.
Note
•
Indicates that you should take special notice.
To obtain a free printed copy of the user documentation, contact Natus Medical Denmark ApS (www.natus.com).
3
Unpacking
1. Unpack the device carefully.
When you unpack the device and accessories, keep the packing material in which they were delivered. If you need to
send the device in for service, the original packing material will protect against damage during transport.
2. Visually inspect the equipment for possible damage.
If damage has occurred, do not put the device into operation. Contact your local distributor for assistance.
3. Check with the packing list to make sure that you have received all necessary parts and accessories. If your package is
incomplete, contact your local distributor.
4. Check the Calibration Certificate to make sure that the transducers (headphones and bone conductor) are the correct
ones, and that they comply with the ordered calibration standards.
4
Installation
Install Otosuite on the PC before you connect Madsen A450 to the PC.
For Otosuite installation instructions, see the Otosuite Installation Guide, on the Otosuite installation medium.
Madsen A450 is fully assembled on delivery, and you simply have to connect the cables.
Warning
•
To connect Madsen A450 to the PC, use the supplied USB cable. The cable length must not
exceed 3 m (approx. 10 feet).
Desktop or wall-mount installation
You can place Madsen A450 on the desktop or mount it on the wall.