background image

Calibration

In the calibration procedure, the patient is asked to switch their gaze 
between the two dots that appear when the lasers are on. If the 
patient cannot be calibrated, click 

Default

 to use the calibration 

default values.

Caution:

 During this procedure both lasers will turn on.  

Do not look directly at the lasers.

1.  Click 

Run

2.  Ask the patient to face the fixation dot and hold their head still.
3.  Ask the patient to look right at the laser beam dot.  

The calibration values are automatically saved.  

Caution:

 Once calibration has been performed, it is recommended 

not to move the goggles.

Calibration Check

1.  Ask the patient to stare at the fixation dot and move the head 

side to side through a small angle (about 10 degrees).

2.  Check that the eye and lateral head trace overlay.
3.  If the eye and lateral head trace do not overlay, you need to 

recalibrate or review the patient’s history.

4.  Once the calibration has been checked and you are satisfied with 

the result, click 

Accept

.

The Collection window opens and you are ready to begin testing.

Head Position Feedback

Clinically, for dynamic positional testing and for repositioning 
treatment, positioning the head properly is essential to acquiring 
the proper diagnosis and successful treatment. Data from the head 
sensor in the goggles tracks head movement. The head movement 
seen in the software corresponds directly to the patient’s head 
movement. This assists the tester to position the head in the various 
positions unique for each dynamic positional test or repositioning 
maneuver.

•  The semicircular canal(s) being tested or treated is highlighted in 

green.

•  As the test or maneuver is performed, the blue dot moves with 

the head.

•  According to the test or treatment maneuver being performed, 

each position is numbered and represented by a blue line.  Move 
the patient’s head aligning the blue dot with each successive 
numbered blue line.

Positional Data Collection

1.   Perform the dynamic positional test or a repositioning maneuver.
2.   Ask the patient to keep their eyes wide open and perform the 

test or treatment as described in the Dynamic Positional or 
Repositioning Training video.

3.  Click the Start button in the software or use a presentation 

remote to start/stop the test and center the head for the head 
position feedback.

A separately purchased presentation remote can be used with the 
Otosuite

®

 Vestibular application. An external monitor displays the 

eye video, real time SPV, and elapsed time separately from the laptop 
making viewing patient response from multiple positions in the room 
possible.

Consult the Reference Manual in the software for more information.

©2021 Natus Medical Incorporated. All Rights Reserved. All product names appearing on this document 

are trademarks or registered trademarks owned, licensed to, promoted or distributed by Natus Medical 

Incorporated, its subsidiaries or affiliates.  06/2021. 7-50-1133-EN Rev04. 

7-50-11303-EN

Healthcare solutions with one thing in mind. You.

natus.com

Natus Medical Incorporated

Positional 

Quick Guide

ICS

®

 Impulse

Отзывы: