OLYMPIC BRAINZ MONITOR 

 QUICK START 

Set up the device at the bedside and connect power 

Turn ON

 using the 

BLUE BUTTON

 

on the front bottom right side of monitor and wait for system to 

power up to the main display 

 

 

Hang the Data Acquisition box from the basket or handle on incubator / warmer 

Prepare for Electrode Application

 

 you will need the following: 

 

2 subdermal needle electrodes 

 

1 hydrogel electrode (for ground) 

 

Electrode positioning aid and Skin Marker 

 

Alcohol swabs for skin cleaning 

 

Tape to secure electrodes 

 

Comb/toothbrush to part the hair if needed 

 

NuPrep and gauze for cleaning ground electrode site 

Measure the infants head keeping the positioning aid vertical and parallel 
to the face.  Align it so that the letter at the ear tragus is the same as the 
letter at the sagittal suture.  The forward edge of the positioning aid should 
touch the tragus.  Mark the two sensor sites, one on each side of the 
positioning aid, at the end of the arrows (P3/P4 position) 

 

Clean electrode site with alcohol swab and insert the needle 
electrode sub dermally.   

 

Place one electrode at a time and secure 

 

For ground electrode, select a site with minimal hair on shoulder, neck or behind the ear 

 

Clean site with NuPrep, remove NuPrep with damp gauze and firmly place electrode 

 
Use the head graphic to connect the sensors into the appropriate 
receptacles on the data acquisition box. The left posterior sensor goes into 
P3, the right posterior sensor goes into P4 and the ground sensor plugs into 
the COMMON receptacle in the middle 
 
 

 

 

 

 

 