64
Setting the High ICP Alarm Limit
Press the
Alarm
tab to specify settings for controlling the physiological alarm. This panel includes arrows for
specifying the high ICP alarm limit value, turning the alarm on and off, and restoring the high ICP alarm limit from a
customized setting to the factory default setting (20 mmHg).
The information on this panel includes:
Item
Description
1
Alarm On
Activates the physiological alarm settings. If this option is selected, the monitor will sound an
alarm if the patient’s displayed Mean ICP exceeds the specified high ICP alarm limit for more
than 5 seconds. By default, this option is selected and set to 20 mmHg.
2
Alarm Off
Deactivates the physiological alarm settings. If this option is selected, the monitor will disable the
high ICP alarm and prevent a high ICP alarm from sounding. In this mode, the monitor will
display
ICP Alarm Off
in the status bar. Note that technical alarms cannot be disabled with this
option.
3
High ICP Alarm Limit
Specifies the maximum ICP value that the monitor will tolerate before activating the alarm. To
specify this limit, press the arrows to a value between -10 mmHg and +125 mmHg in 1 mmHg
increments.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...