Bio-logic
®
AuDX PRO FLEX
®
QUICK START GUIDE
3
Connect the probe to the red connector and the air cable to the pump connector. From the start menu, select the
Immittance:
Tymp+Ref button
. Select the test ear by pressing either the Right or Left button. Cartoon mode will show an airplane on
the screen during the tympanometry test. For the diagnostic module, connect a second transducer to the blue connector for
Contralateral Reflex testing.
The system will first check the probe fit and wait for
the pressure build up. The LED light on the probe will
pulse at a steady pace to indicate a correct fit. If the
ear tip fit is not correct, the LED light will blink at a
fast pace and an error message prompting you to
retry or cancel the test will appear on the screen.
Once a proper seal has been obtained, the test will proceed. A tympanometry test will be performed, followed by a reflex
measurement.
Tympanometry result
Reflex threshold found
Reflex test in progress
Overall Reflex test result
Select the appropriate ear tip size and place it on the probe tip.
Insert the probe in the ear making sure that the eartip fits tightly into the ear canal.
Testing: Tympanometry
Tympanometry test screen
Air Pump
Second transducer
Probe or
second transducer