background image

 

 

 
MAINTENANCE 

 

WARNING:  For your own safety, turn switch off and remove plug from power source before maintaining 
your microscope.  If the power cord is worn, cut or damaged in any way, have it replaced immediately to 
avoid shock or fire hazard. 

 
1.   OPTICAL MAINTENANCE 
 

A. 

Do not attempt to disassemble any lens component.  Consult an expert technical service company 
when repairs not covered by these instructions are required. 
 

B. 

Prior to cleaning any lens surface, brush dust or dirt off lens surfaces using a camel hair brush.  Or use 
air to blow dust and lint off surfaces.  Use of compressed air in a can,  available at any computer 
supply store, is a good source of clean air. 
 

C.  Cleaning eyepiece lenses. 
 

Do not remove eyepiece from eyepiece tube.  Clean only the outer surface.  Breath on lens to dampen 
surface, then wipe with lens paper.  Do not wipe lens surface while dry as lenses are scratched very 
easily.  Wipe in a circular motion from center to outer edges. 
 

D.  Cleaning objective lenses. 
 

Your microscope is supplied with a clear glass objective lens protector; do not remove this lens 
protector from microscope.  Clean glass element only.  Using a cotton swab saturated with distilled 
water, clean front surface.  If specimen material of any kind is evident, use a cotton swab dipped in a 
small amount of distilled water or Windex to clean all foreign material from glass surface.   

 
2.  MECHANICAL MAINTENANCE 
 

The only mechanical adjustment the microscope might require is the tension of the focusing mechanism.  
This has been adjusted at the factory, but over the course of time it may loosen and cause the head of 
the microscope to slip downward on the focusing block. 
 
The tension adjustment collar is located between arm and focus knob on left side of microscope.  With a 
jewelers type screwdriver, loosen slotted set screw located on knurled surface of the tension adjustment 
collar.  Turn collar clockwise to tighten tension, counter-clockwise to loosen tension.  After adjusting, 
tighten the set screw to lock collar in place. 
 
NOTE:  It is recommended that you leave the tension as loose as possible for ease of focusing, yet not 
so loose that it permits the head of microscope to drift downward from its own weight and cause the 
microscope to “drift” out of focus. 

 
3.  ELECTRICAL MAINTENANCE  
 

The extent of electrical maintenance, by other than a qualified technician, should be bulb or fuse 
replacement.  BE CERTAIN TO TURN SWITCHES OFF AND REMOVE PLUG FROM POWER SOURCE 
OUTLET BEFORE CHANGING BULBS. 

 

A.  To replace top bulb... remove black top light housing by rotating in a counter-clockwise direction.  

Remove lamp using a clean cloth by depressing and rotating lamp.  Install new lamp using a clean cloth, 
gently depress into socket and turn clockwise. Replace light housing. 

 

Содержание 409TBL Binocular Stereo

Страница 1: ... Tri County Parkway Schertz Texas 78154 Phone 210 590 9010 Fax 210 590 1104 INSTRUCTIONS FOR STEREOSCOPIC MICROSCOPES 409TBL Binocular Stereo 410TBL Binocular Stereo 411TBL Binocular Stereo Copyright 1 2 01 National Optical Scientific Instruments Inc ...

Страница 2: ... Stage plate locking screw Stage plate Fuse Top light incidental illumination Arm Focusing knob tension adjustment collar behind knob on this side Head Diopter adjustment Eyepiece locking screw Main power switch Stage clips Rubber eyepiece shield Slotted head retaining screw located on side Glass objective lens protector ...

Страница 3: ...hanging stage plate G Dustcover 2 Examine packing material before you discard it Retain the styrofoam container in case you need to transport store or return the microscope for service If it becomes necessary to ship the microscope for any reason pack it in the styrofoam container and then pack the styrofoam in another corrugated shipping container for optimum protection Use of the styrofoam alone...

Страница 4: ...l help to secure them against loss B Install rubber eyepiece shields by slipping over top of eyepieces flared side outward OPERATION 1 ILLUMINATION A Before operating microscope make certain that the main voltage of your microscope corresponds to the voltage of your power outlet either 120v or 220v Insert microscope plug into matching voltage outlet B The microscope is furnished with two stage pla...

Страница 5: ...en slide cover slip up on center of stage plate C Turn focusing knobs until object being viewed is in focus D Adjust diopter for difference in eyesight Look into right eyepiece with right eye Turn focusing knobs until sharp image is obtained in right side of microscope Look into left eyepiece with left eye Adjust sharpness of image by turning knurled diopter adjustment located on left eyepiece tub...

Страница 6: ...in a small amount of distilled water or Windex to clean all foreign material from glass surface 2 MECHANICAL MAINTENANCE The only mechanical adjustment the microscope might require is the tension of the focusing mechanism This has been adjusted at the factory but over the course of time it may loosen and cause the head of the microscope to slip downward on the focusing block The tension adjustment...

Страница 7: ...ower cord not connected Plug into outlet Lamp burned out Replace lamp Fuse blown Replace fuse Image does not remain in focus Head of microscope drops from its own weight Adjust tension control Poor resolution image not sharp Clear glass objective lens protector dirty Clean glass protector Eyepiece lens dirty Clean eyepiece lenses Spots in field of view Eyepiece lens dirty Clean eyepiece lenses Spo...

Страница 8: ...o be free from defects in material and workmanship under normal use and service for 90 days from date of purchase Other than set forth above National hereby disclaims all warranties express or implied of fitness for a particular purpose While it is not necessary to register your purchase of a microscope any camera or software must be registered by completing and mailing the warranty registration c...

Отзывы: