NI WSN-3226 Calibration Procedure
|
© National Instruments
|
3
•
Maintain an ambient temperature of 23 °C ± 5 °C.
•
Keep relative humidity between 10% and 90%.
•
Allow a warm-up time for all the instruments and equipment, according to the
manufacturers’ instructions.
Initial Setup
Refer to the
NI Wireless Sensor Network Devices Getting Started Guide
for information about
how to install the software and hardware and how to configure the device in Measurement &
Automation Explorer (MAX).
Verification
The following performance verification procedures describe the sequence of operation and test
points required to verify the NI WSN-3226 and assume that adequate traceable uncertainties are
available for the calibration references.
Voltage Accuracy Verification
Complete the following procedure to determine the As-Found status of the NI WSN-3226.
1.
Connect the calibrator to the NI WSN-3226 as shown in Figure 1.
Figure 1.
NI WSN-3226 Voltage Connections
2.
Configure a project for the WSN system in LabVIEW.
3.
Right-click the NI WSN-3226 in the Project Explorer and select
Properties
to launch the
Properties dialog box.
C
a
li
b
r
a
tor
NI W
S
N-
3
226
COM
3
AI
3
COM 2
AI 2
COM 1
AI 1
COM 0
AI 0
LO
HI