NI VirtualBench Calibration Procedure
|
© National Instruments
|
63
8.
Call
PS Read Output
with the following parameter:
•
Channel
=
ps/+6V
9.
Subtract the DMM reading from step 7 from the DC power supply Actual Current Level
reading in step 8. Verify that calculation falls between the limits in Table 22 or 23.
10. Repeat steps 5 through 9 for each verification point on this channel.
11. Call
PS Reset Instrument
.
12. Disconnect all connections and connect C25 V to Amps HI, and Common Floating
GND to Amps LO as shown in Figure 33. Connect GND and Common Floating GND with
a jumper.