National Instruments SCXI-1382 Скачать руководство пользователя страница 4

©

 National Instruments Corporation

3

SCXI-1382 Battery Pack Installation Guide

Battery Pack Safety Information

Note

Refer to the 

Read Me First: Safety and Radio-Frequency Interference

 document 

before removing equipment covers or connecting or disconnecting any power wires or 
signal wires.

Cautions

The battery pack contains a sealed lead-acid non-wet, nonspillable battery.

Do 

not

 ship the battery pack without proper packaging or if it is damaged. Retain the 

original packaging for future shipping.

Do 

not

 carry the battery pack onto a passenger plane. You must ship the battery pack as 

cargo.

You must recycle or properly dispose of the battery. In the United States, call 
1-800-RE-USE-Pb (1-800-738-7372) for more information.

Never

 expose the battery to fire or short-circuit the battery.

Do 

not

 use the battery pack if it appears damaged.

Always use the supplied charger to charge the battery pack. Follow the documentation 
instructions. You can obtain NI documents at 

ni.com

 or by contacting an NI sales 

representative.

Charging the Battery Pack

The SCXI-1382 battery pack requires 12 to 14 VDC and at least 3.7 A of 
power for at least 12 to 14 hours to reach a full charge. Charge the battery 
pack as soon as possible after use. If the battery is fully charged, it can 
power the SCXI-1000DC chassis for more than five hours. However, 
NI recommends that you charge the battery after every three hours of use 
to extend the life of the battery pack.

Note

NI recommends using the SCXI-1383 power supply/float charger to charge the 

SCXI-1382 battery pack. Refer to the 

SCXI-1383 Power Supply/Float Charger 

Installation Guide

 for charging instructions.

Caution

The battery pack contains a sealed lead-acid battery that contains toxic material 

(lead) and corrosive fluid (sulfuric acid). Charging can produce explosive gases. Do 

not

 

charge the battery in gas-tight enclosures. Charge the battery in a well ventilated area away 
from sparks, flames, or smoke. Do 

not

 short the battery pack terminals, as this can cause 

an explosion or fire. Do 

not

 store the battery pack in a discharged state. Do 

not

 puncture, 

disassemble, mutilate, or incinerate the battery pack. Recycle the battery or dispose of the 

Содержание SCXI-1382

Страница 1: ...erutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos In st...

Страница 2: ...chassis The SCXI 1383 is a constant voltage current limited charger There are two versions of the SCXI 1382 battery pack kit One kit requires an SCXI 1383 and one does not If the kit contains an SCXI...

Страница 3: ...82 battery pack Mounting bracket for the SCXI 1382 battery pack attached to the battery pack Screws 8 32 5 16 in flathead Screw 6 32 1 4 in flathead SCXI 1382 Battery Pack Installation Guide Read Me F...

Страница 4: ...g an NI sales representative Charging the Battery Pack The SCXI 1382 battery pack requires 12 to 14 VDC and at least 3 7 A of power for at least 12 to 14 hours to reach a full charge Charge the batter...

Страница 5: ...1 2 1 4 N m 10 5 12 3 lb in to the screws to secure them 4 Slide the battery pack assembly onto the mounting bracket and reinsert the four 8 32 7 16 in screws Apply a torque of 1 2 1 4 N m 10 5 12 3...

Страница 6: ...ack Installation Guide Figure 1 Connecting the SCXI 1382 to the SCXI 1000DC Chassis 1 SCXI 1000DC Chassis 2 8 32 x 7 16 in Screws 3 Mounting Bracket 4 8 32 x 5 16 in Screws 5 SCXI 1382 Battery Pack 6...

Страница 7: ...ollow steps 1 to 6 in the Connecting the SCXI 1382 to the SCXI 1000DC Chassis section 2 Connect the green DC power connector of the SCXI 1383 power supply float charger P2 to its corresponding connect...

Страница 8: ...are typical at 25 C unless otherwise specified Note Refer to the SCXI 1383 Power Supply Float Charger Installation Guide for its specifications Battery Pack Battery type Sealed lead acid Voltage 12 V...

Страница 9: ...standards of safety for electrical equipment for measurement control and laboratory use IEC 61010 1 EN 61010 1 UL 3111 1 UL 61010B 1 CAN CSA C22 2 No 1010 1 Note For UL and other safety certification...

Страница 10: ...Directives as amended for CE marking as follows Low Voltage Directive 73 23 EEC safety Electromagnetic 89 336 EEC Compatibility Directive EMC Note Refer to the Declaration of Conformity DoC for this p...

Страница 11: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: