background image

5

MIO Board

Figure 4.  Removing the Key Inserts

e. Identify the proper replacement bracket from your kit.  If you have an NB Series board,

use the bracket marked 745402-01; if you have an AT Series board, use the bracket

marked 745403-01.  The bracket that you use should look like the one you remove,

except that the new one has an extra tab that attaches to the grounding strap.

f. Install the new bracket by aligning the bracket with the holes in your MIO board and

screw in the two mounting screws.

6. Install your MIO board in a slot in your computer, following the instructions in your MIO

board user manual.

7. Connect the 68-pin connector (the end without the grounding strap) of the SCXI-1345

cable to the 68-pin connector of the SCXI-1345 adapter board as shown in Figure 5.

Adapter

Board

Shielded Cable

Figure 5.  Cable to Adapter Board Connection

Содержание SCXI-1345

Страница 1: ...chassis and your MIO board You can use the SCXI 1345 with the following MIO boards PC XT PC AT Macintosh AT MIO 16 NB MIO 16 AT MIO 16F 5 NB MIO 16X AT MIO 16X The SCXI 1345 shielded cable consists of...

Страница 2: ...Acquisition Board Figure 1 Connecting the SCXI 1345 to Your MIO Board and SCXI Module Perform the following steps to install the SCXI 1345 cable 1 Turn off the power to your computer and the SCXI cha...

Страница 3: ...ounding strap tab a Remove your MIO board from your computer b Unscrew and remove the two screws that attach the bracket to the MIO board c Remove the bracket from the board See Figure 2 for NB board...

Страница 4: ...ompatible with the 50 position connector on the SCXI 1345 you must remove them The 50 position SCXI 1345 connector has a center polarization bump that prevents you from inserting the connector upside...

Страница 5: ...an extra tab that attaches to the grounding strap f Install the new bracket by aligning the bracket with the holes in your MIO board and screw in the two mounting screws 6 Install your MIO board in a...

Страница 6: ...of the ground strap onto the tab of the MIO bracket 10 Secure the cable to a fixed object with the tie wraps to relieve the strain on the cable Strain relief is needed because the SCXI 1345 shielded c...

Страница 7: ...ase the ground current flowing between the SCXI system and the computer thus increasing the noise in some applications Laboratory tests with the grounding strap connected produced less noise but you c...

Отзывы: