background image

©

 National Instruments Corporation

9

NI SCXI-1332 Installation Instructions

Specifications

Maximum Voltage

Channel-to-ground ................................. 300 V, CAT II

Caution

This module is rated for Measurement Category II and intended to carry signal 

voltages no greater than 300 V. This module features 1390 V

rms

 isolation between input 

signals and the chassis as verified by a dielectric withstand test, 1 minute maximum. Do

not 

use this module for connection to signals or for measurements within Categories III or IV. 
Refer to the 

Read Me First: Safety and Radio-Frequency Interference 

document for more 

information about measurement categories.

Maximum Current

Maximum current
(per channel) .......................................... 2 ADC, 2 AAC

Environment

The NI SCXI-1332 is intended for indoor use only.

Operating temperature............................ 0 °C to 50 °C

Storage temperature ............................... –20 °C to 70 °C

Relative humidity ................................... 10% to 90% RH, noncondensing

Pollution Degree .................................... 2

Approved at altitudes up to 2,000 m.

Safety

This product is designed to meet the requirements of the following 
standards of safety for electrical equipment for measurement, control, 
and laboratory use:

IEC 61010-1, EN 61010-1

UL 61010-1

CAN/CSA-C22.2 No. 61010-1

Note

For UL and other safety certifications, refer to the product label or visit 

ni.com/certification

, search by model or product line, and click the appropriate link 

in the Certification column.

Содержание SCXI-1332

Страница 1: ...ons Conventions The following conventions are used in this guide The symbol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup Options directs yo...

Страница 2: ...andling the terminal block take the following precautions Caution Never touch the exposed pins of connectors Ground yourself using a grounding strap or by touching a grounded object Touch the antistat...

Страница 3: ...nd of the wire 2 Remove the top cover screw 3 Unsnap and remove the top cover 4 Loosen the two strain relief screws on the strain relief bar 5 Run the signal wires through the strain relief opening 6...

Страница 4: ...NI SCXI 1332 Installation Instructions 4 ni com Figure 1 SCXI 1332 Top Cover Diagram 1 Top Cover Screw 2 Top Cover 2 1...

Страница 5: ...al blocks If you are not expanding the number of columns of the matrix skip to the 6 Install the Terminal Block step For more information about expanding matrices refer to the NI Switches Help 1 Colum...

Страница 6: ...one end of the matrix expansion cable to a row connector in one of the terminal blocks 2 Connect the other end of the matrix expansion cable to a row connector in another terminal block Figure 3 Expa...

Страница 7: ...e following steps 1 Connect one end of a matrix expansion cable to a column connector in one of the terminal blocks 2 Connect one end of a second matrix expansion cable to another column connector in...

Страница 8: ...the following steps 1 Plug the SCXI 1332 onto the front connector on the SCXI 1127 1128 2 Tighten the top and bottom thumbscrews on the terminal block rear panel to hold it securely in place Figure 5...

Страница 9: ...document for more information about measurement categories Maximum Current Maximum current per channel 2 ADC 2 AAC Environment The NI SCXI 1332 is intended for indoor use only Operating temperature 0...

Страница 10: ...liance This product meets the essential requirements of applicable European Directives as amended for CE marking as follows Low Voltage Directive safety 73 23 EEC Electromagnetic Compatibility Directi...

Страница 11: ...al Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the a...

Отзывы: