Chapter 5
Calibration
5-2
www.natinst.com
6.
Connect the amplifier output to the DMM. Make sure that the DMM
can achieve the accuracy and resolution you need.
7.
Adjust the output potentiometer of the desired channel until the output
is 0 ± 3 mV.
8.
Set the DMM range to the smallest range that can measure ±6 mV.
9.
Set the amplifier gain to 1,000.
10. Adjust the input potentiometer of the desired channel until the output
is 0 ± 6 mV.
11. Go to the next channel.
You may seal the potentiometers after calibration with antisabotage lacquer
to avoid tampering with the calibration.
Table 5-1 lists the potentiometer reference designators that correspond to
each channel.
Table 5-1.
Calibration Potentiometers Reference Designators
Input
Channel
Number
Amplifier Channel
Input Null
Output Null
0
R08
R24
1
R10
R25
2
R12
R26
3
R14
R27
4
R16
R28
5
R18
R29
6
R20
R30
7
R21
R31