SCXI-1120/D Calibration Procedure
4
ni.com
Setting Up the Module
Refer to Figures 1 and 2 while performing the following steps to set up the
SCXI-1120/D module for verification:
1.
Remove the grounding screw from the module.
2.
Remove the cover on the module to access the potentiometers.
Figure 1.
Grounding Screw and Module Cover Removal
3.
Remove the side plate of the SCXI chassis.
4.
Install the SCXI-1120/D into slot 4 of the SCXI chassis.
1
Grounding screw
2
Module cover
1
2