SCXI-1104 Calibration Procedure
•
DAQGain
—1 for 16-bit devices, –1 for 12-bit devices
•
Tbgain
—1.0
•
volt1
and
volt2
—The first and second voltage readings for the
gain values that you are about to save on the SCXI-1104
•
binary1
and
binary2
—The first and second binary readings for
the gain values that you are about to save on the SCXI-1104
•
calConst1
and
calConst2
—These are return values.
2.
Repeat step 1, changing
calibrationArea
to 1.
3.
Repeat step 1, changing
calibrationArea
to 3.
4.
Repeat steps 1 through 3 for the next channel.
You have now adjusted the calibration constants.
Verifying Adjusted Values
After completing the adjustments, it is important to verify the adjusted
values by repeating the steps listed in the
Verifying the Operation of Your
section of this document. Verifying the adjusted values ensures
your module is now operating within its specifications.