Using the SCC-RTD01

Note

If you have an RTD that is not supported in Measurement & Automation 

Explorer (MAX), Read Voltage Data and scale it in your ADE.

Converting Voltage Measurements to Temperature Measurements

If you use MAX to configure the SCC-RTD01, the reading you get from 
the E/M Series DAQ device is properly scaled. Otherwise, you must scale 
the readings and convert the voltage measurement to temperature by 
performing the following steps to use the standard equations for platinum 
RTDs:

1.

Measure the RTD voltage.

a.

Read the RTD channel on the E/M Series DAQ device and call the 
value 

V

ESERIES

[CH(

X

)].

b.

Calculate the RTD voltage using the following formula:

where

V

RTD

 is the SCC-RTD01 input voltage.

V

ESERIES

 is the E/M Series DAQ device voltage.

2.

Use polynomial expressions or a conversion table provided by the RTD 
vendor to convert the RTD voltage to temperature.

Note

NI programming environments include RTD conversion utilities that implement the 

voltage-to-temperature conversions. Refer to the ADE documentation for more 
information on these utilities.

Although the RTD resistance-versus-temperature curve is relatively 
linear, accurately converting resistance to temperature requires curve 
fitting. The Callendar-Van Dusen equation is commonly used to 
approximate the RTD curve:

R

t

 = 

R

0

[1 + A

t

 + B

t

2

 + C(

t

 – 100)

3

]

where

R

t

 is the resistance of the RTD at temperature 

t.

R

0

 is the resistance of the RTD at 0 

°

C.

A, B, and C are the Callendar-Van Dusen coefficients shown in 
Table 1.

t

 is the temperature in 

°

C.

V

RTD

V

ESERIES

25

---------------------

=

Содержание SCC-RTD01

Страница 1: ...rent source for RTDs Conventions The following conventions are used in this guide The symbol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup O...

Страница 2: ...g examples and syntax examples This font is also used for the proper names of disk drives paths directories programs subprograms subroutines device names functions operations variables filenames and e...

Страница 3: ...re scaling Device Specific Information Note For general SCC module installation and signal connection information and information about the SC 2345 or SC 2350 carrier refer to the SCC Quick Start Guid...

Страница 4: ...measuring the voltage across RTDs Pins 1 and 2 form a differential channel routed to E M Series DAQ device channel X 8 and pins 3 and 4 form a second differential channel routed to E M Series DAQ devi...

Страница 5: ...hannel Red Red Black Black RTD IEX 1 mA E M Series AI X E M Series AI X 8 6 5 4 3 2 1 SCC RTD01 2 Wire 1 Channel Red Black RTD IEX 1 mA E M Series AI X E M Series AI X 8 6 5 4 3 2 1 SCC RTD01 3 Wire 1...

Страница 6: ...ions can introduce measurement errors For information about configuring the SCC RTD01 module using NI DAQmx refer to the SCC Quick Start Guide IEX 1 mA 6 5 4 3 2 1 SCC RTD01 4 Wire 2 Channels Red Red...

Страница 7: ...te the RTD voltage using the following formula where VRTD is the SCC RTD01 input voltage VESERIES is the E M Series DAQ device voltage 2 Use polynomial expressions or a conversion table provided by th...

Страница 8: ...sured RTD voltage IEX is the excitation current Most platinum RTD curves conform to one of the following three standardized curves DIN 43760 standard U S Industrial or American standard International...

Страница 9: ...M in parallel with 4 7 nF Powered off 20 k min Overload 20 k min Filter type 3 pole Butterworth lowpass filter 3 dB cutoff frequency 30 Hz System noise 4 5 Vrms referred to the input RTI 1 Transfer Ch...

Страница 10: ...ut regulation loss 24 V Drift 127 ppm C Power Requirement Analog power 135 mW max 15 V 4 5 mA max 15 V 4 5 mA max Digital power 153 mW max 5 V 30 6 mA max Physical Figure 3 SCC RTD01 Dimensions Weight...

Страница 11: ...or electrical equipment for measurement control and laboratory use IEC 61010 1 EN 61010 1 UL 61010 1 CAN CSA C22 2 No 61010 1 Note For UL and other safety certifications refer to the product label or...

Страница 12: ...nd click the appropriate link in the Certification column Waste Electrical and Electronic Equipment WEEE EU Customers At the end of their life cycle all products must be sent to a WEEE recycling cente...

Страница 13: ...appropriate location Help Patents in your software the patents txt file on your CD or ni com patents 2002 2006 National Instruments Corporation All rights reserved 371072C 01 Jul06 AI X 8 A GND is the...

Отзывы: