Constant voltage control using an Vext or Rext (CF06)
•
External control setting for turning output on and off (CF09)
•
Voltage and current setting limits (CF15 and CF14)
CF01 Master-slave Parallel Operation Parameter
Sets the total number of units (master unit and slave units) in master-slave parallel operation.
Specify “SLAV” for the units that you want to use as slaves. This parameter is applied when you
turn the POWER switch off and on.
Changing this setting may affect the settings of the master unit or the slave units that are
connected in parallel.
CF02 Power-on Status Parameter
Sets the condition that the RMX programmable power supply will be in when the power is turned
on. This setting is invalid when you are using an external contact to turn the output off. This
parameter only applies when you turn the POWER switch off and on.
You can only specify this parameter when the RMX programmable power supply is being used
as a master unit.
Value
Parameter Description
rSt
Resets the panel settings
Value
Parameter Description
SLAV
Use the RMX as a slave unit
1 to 4
Set the total number of units (1 to 4; 1 indicates independent
operation–factory default setting).
Value
Parameter Description
SAFE
The settings that were in use immediately before the POWER
switch was turned off are used. The output is off (factory default
setting).
Auto
The settings that were in use immediately before the POWER
switch was turned off are used. This includes whether output was
on or off.
ForC
The settings that were in use immediately before the POWER
switch was turned off are used. The output is turned on.