PXIe-6595 Kit Contents
The following items are included in the device kit:
•
PXIe-6595
•
PXIe-6595 Safety, Environmental, and Regulatory Information
document
Preparing the Environment
Ensure the environment in which you are using the PXIe-6595 meets the following
specifications.
Operating environment
Ambient temperature range
0 °C to 55 °C
1
Relative humidity range
10% to 90%, noncondensing
Maximum altitude
2,000 m (800 mbar) (at 25 °C ambient
temperature)
Pollution Degree
2
Indoor use only.
Note
For complete specifications, refer to the specifications document for your
.
Installing the PXIe-6595
Notice
To prevent damage to the PXIe-6595 caused by ESD or contamination,
handle the module using the edges or the metal bracket.
1.
Ensure the AC power source is connected to the chassis before installing the module.
The AC power cord grounds the chassis and protects it from electrical damage while you
install the module.
2.
Power off the chassis.
3.
Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.
4.
Remove the black plastic covers from all the captive screws on the module front panel.
5.
Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.
1
The PXIe-6595 requires a chassis with slot cooling capacity ≥58 W. Not all chassis with slot
cooling capacity ≥58 W can achieve this ambient temperature range. Refer to the
for specifications to determine the ambient temperature ranges your chassis can achieve.
2
|
ni.com
|
PXIe-6595 Getting Started Guide