Table 2. Access LED Status Indicator

Color

Indications

 

 

Off

Instrument not yet functional.

Amber

Instrument being accessed by software.

Green

Instrument ready to be programmed.

Use the following table to determine the PXIe-6570 state using the Active LED status
indicator.

Table 3. Active LED Status Indicator

Color

Indications

 

 

Off

Instrument is not waiting for a trigger, not bursting a pattern, and not experiencing
an error. During PPMU operations, the Active LED remains off.

Amber Instrument awaiting Start trigger.

Green Instrument received Start trigger and is bursting a pattern.

Red

Instrument error condition.
Possible error conditions:

Self-test failed on the instrument. This error condition will persist until the
instrument is reset or self-test succeeds.

Self-calibration or external calibration adjustment failed on the instrument. This
error condition will persist until the instrument is reset or calibration succeeds.

The instrument has detected an unlocked condition on a PLL that might result
in incorrect data. To clear this error condition, reset the instrument
programmatically or through Measurement & Automation Explorer (MAX).

The instrument has been disabled because it exceeded its overall power limit.
To re-enable the instrument, reset it programmatically or through MAX.

The instrument has been disabled because it exceeded its overall temperature
limit. To re-enable the instrument, cool the instrument to an acceptable range
and resolve the environmental condition that caused the shutdown. Reset the
instrument programmatically or through MAX, or power cycle the instrument.

Connecting to a Device Under Test

To connect a device under test (DUT) to the PXIe-6570, you can create a custom device
interface board to access the signals of the DUT. For more information, refer to the 

Interfacing

to the Digital Pattern Instrument or Digital Waveform Instrument using the VHDCI Connector

PXIe-6570 Getting Started Guide

  | 

© National Instruments

  | 

9

Содержание PXIe-6570

Страница 1: ...quirements 2 Unpacking the Kit 2 Kit Contents 3 Preparing the Environment 3 Installing the Software 3 Digital Pattern Editor 4 NI Digital Pattern Driver 4 Associated Drivers 5 Installing the PXIe 6570...

Страница 2: ...ielded cables or accessories unless they are installed in a shielded enclosure with properly designed and shielded input output ports and connected to the product using a shielded cable If unshielded...

Страница 3: ...ver and Digital Pattern Editor media NI part number 785249 03 separately Preparing the Environment Ensure that the environment in which you are using the PXIe 6570 meets the following specifications O...

Страница 4: ...tate of the digital pattern instrument The Digital Pattern Editor is installed when you install NI Digital Pattern Driver For more information refer to the Digital Pattern Help To access Digital Patte...

Страница 5: ...g the PXIe 6674T with digital pattern instruments refer to the Digital Pattern Help Visit ni com downloads to download NI Sync Installing the PXIe 6570 Caution To prevent damage to the PXIe 6570 cause...

Страница 6: ...ure 2 PXIe 6570 Installation 4 3 2 1 1 PXI Express Chassis 2 Ejector Handle in Down Position 3 NI PXI Express Two Slot Module 4 Captive Screw 8 Secure the module front panel to the chassis using the f...

Страница 7: ...68 67 66 65 64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 38 37 36 35 GND CAL FORCE GND DIO 0 CAL GND DGS DIO 2 GND DIO 4 RESERVED DIO 6 GND DIO 8 GND DIO 10 GND DIO 1...

Страница 8: ...und Optional DGS for improved accuracy at higher currents in some configurations CAL GND Analog Resource for external calibration 67 CAL FORCE Analog Resource for external calibration 30 60 RESERVED N...

Страница 9: ...il the instrument is reset or calibration succeeds The instrument has detected an unlocked condition on a PLL that might result in incorrect data To clear this error condition reset the instrument pro...

Страница 10: ...operly designed and shielded input output ports and connected to the product using a shielded cable If unshielded cables or accessories are not properly installed and shielded the EMC specifications f...

Страница 11: ...ier MAX assigns to the hardware Use this identifier when programming the PXIe 6570 5 Self test the hardware by selecting the item in the configuration tree and clicking Self Test in the MAX toolbar Th...

Страница 12: ...tions using the PXIe 6570 Digital Pattern Instrument Examples To locate programming and operating examples refer to the following table Table 6 NI Digital Pattern Driver Example Locations Application...

Страница 13: ...nterfaces 2 Expand the Chassis tree to see the list of installed hardware and press F5 to refresh the list 3 If the module is still not listed power off the system ensure that all hardware is correctl...

Страница 14: ...is powered on Applying external signals while the module is powered off may cause damage Disconnect any signals from the PXI Express module front panels 2 Remove any front panel connections from the...

Страница 15: ...ments National Instruments NET Support Digital Pattern Help Worldwide Support and Services The NI website is your complete resource for technical support At ni com support you have access to everythin...

Страница 16: ...oducts technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the National Instruments Patent Notice at ni com patents You can find informatio...

Отзывы: