NI PXIe-5694 Calibration Procedure
|
© National Instruments
|
29
13. Repeat steps 4 through 11 with the new settings and the following changes:
•
When configuring RF source 1 and 2 output power, use a device under test (DUT)
input power of -5 dBm instead of -0 dBm.
•
Record your results using Table 9.
14. Reconfigure the spectrum analyzer for a frequency of 21.4 MHz.
15. Reconfigure the NI 5694, enabling IF conditioning downconversion and disabling step
gain.
Table 9.
TOI-OB Calculation,
DUT Input Power -5 dBm (per Tone), Downconversion Disabled
Device
Instantaneous
Bandwidth (Path)
RF Source 1
Frequency
(MHz)
RF Source 2
Frequency
(MHz)
IF OUT
Power
(dBm)
TOI-OB
(dBm)
20 MHz
223.6
253.6
163.6
133.6
5 MHz
201.1
208.6
186.1
178.6
1.4 MHz
195.7
197.8
191.5
189.4
400 kHz
194.2
194.8
186.1
178.6
110 kHz
193.765
193.93
193.435
193.51