26
|
ni.com
|
NI PXIe-5693 Calibration Procedure
•
Detector type: Auto Peak
•
Average count: 5
•
External reference source: EXT
13. Configure RF source 1 and RF source 2 according to the following settings:
•
Single frequency mode: On
•
Power level: -10 dBm
14. Set the power sensor A settling percentage to 0.1% and program the power meter to trigger
when settled.
15. Set the NI 5693 center frequency according to the first row in Table 7.
16. For each NI 5693 center frequency in Table 7, calculate the following two frequencies:
•
Tone 1 Frequency = NI 5693 Center Fre
1.2 MHz - (700 kHz/2)
•
Tone 2 Frequency = NI 5693 Center Fre
1.2 MHz + (700 kHz/2
Table 7.
In-Band TOI Verification Test Frequencies
Center Frequency (MHz)
Tone Spacing (kHz)
Combiner
27
700
Low Frequency
47
700
Low Frequency
80
700
Low Frequency
130
700
Low Frequency
193
700
Low Frequency
288
700
Low Frequency
463
700
Low Frequency
763
700
High Frequency
1,255
700
High Frequency
1,780
700
High Frequency
2,250
700
High Frequency
2,750
700
High Frequency
3,400
700
High Frequency
4,200
700
High Frequency
5,200
700
High Frequency
6,400
700
High Frequency