PXIe-5673 Calibration Procedure
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5673 meets
published specifications.
•
Verify that the PXIe-5611, PXIe-5450, and PXI-5650/5651/5652 are properly connected as
indicated in the
NI RF Signal Generators Getting Started Guide
before starting verification.
•
Keep cabling as short as possible. Long cables and wires act as antennas, picking up extra
noise that can affect measurements.
•
Verify that all connections, including front panel connections and screws, are secure.
•
Use a torque wrench appropriate for the type of RF connector you are using.
NI recommends using a 0.565 N · m (5 lb · in.) torque wrench for SMA connectors and a
0.90 N · m (8 lb · in.) torque wrench for 3.5 mm connectors.
Power splitter
Aeroflex/Weinschel
1593
Output power level
accuracy
VSWR: 1.25:1
Amplitude tracking:
<0.25 dB
6 dB attenuator
Anritsu 41KB-6
Frequency range: DC
to 12 GHz
VSWR: 1.1:1
SMA(m)-to-SMA(m)
adapter
(H+S)
32_N-SMA-50-1/
11_NE
VSWR: 1.05:1
50
Ω
terminator
(included in the
PXIe-5673 kit)
NI 778353-01
—
—
PXIe-5673 cable
accessory kit
(Matched-length I/Q
semi-rigid SMA cables
(4x)
LO semi-rigid SMA
cable (1)
RF flexible SMA cable
(1x))
NI 780567-01
—
—
SMA torque wrench
(1 N · m)
NI 187106-01
—
—
Table 2.
Recommended Equipment for PXIe-5673 Calibration (Continued)
Equipment
Recommended
Model
Where Used
Minimum
Requirements