Table 5.
NI 5442 AWG Module Front Panel Connectors
Connector
Use
CH 0
Connects to the IF INPUT connector on the NI 5610 front panel.
Output terminal for an IF waveform for upconversion to the desired RF
frequency.
CLK IN
Connects to the REF OUT connector on the NI 5610 module front panel.
Input terminal for the NI 5442 AWG module Reference Clock signal.
PFI 0
Both connectors accept a trigger from an external source to start or step through
signal generation. Both terminals are bidirectional SMB connectors.
PFI 1
Table 6.
NI 5442 AWG Module Front Panel LEDs
LED
Indications
ACCESS Indicates the basic hardware status of the module.
OFF—The module is not yet functional or has detected a problem with a power
rail.
AMBER—The module is being accessed.
Accessed
means that the device setup
registers are being written to in order to control the device or load waveforms.
GREEN—The module is ready to be programmed by NI-RFSG.
ACTIVE Indicates the state of the module.
OFF—The module is not armed or triggered.
AMBER—The module is armed and waiting for a Start Trigger.
GREEN—The module has received a Start Trigger and is generating a waveform.
RED—The module has detected an error state; this state may indicate a PLL lock
failure or a thermal shutdown condition.
Configuring the NI 5672 in MAX
Use Measurement & Automation Explorer (MAX) to configure your National Instruments
hardware. MAX informs other programs about which devices reside in the system and how
they are configured. MAX is automatically installed with NI-RFSG.
1.
Launch MAX.
16
|
ni.com
|
NI PXIe-5672 Getting Started Guide