When there is a change to the system cabling
•
When the system is in an environment where external variables, such as temperature, can
affect measurements
You should also perform a self-calibration if your device has exceeded any of the following
temperature or time limits since the last device self-calibration:
•
IF flatness: ±5 °C and/or 7 days
•
Gain reference: ±5 °C and/or 7 days
•
LO self-calibration: ±10 °C and/or 30 days
•
Digitizer self-calibration: ±5 °C and/or 90 days
Performing a Device Self-Calibration Using the
NI-RFSA SFP
NI recommends you run the self-calibration from the NI-RFSA Soft Front Panel (SFP). You
can also run a self-calibration programmatically using the NI-RFSA API.
1. Launch the NI-RFSA SFP by navigating to
Start
»
All Programs
»
National
Instruments
»
NI-RFSA
»
NI-RFSA Soft Front Panel
.
2. Click
Device/System
»
Calibration
»
Self Calibration
.
3. Select the self-calibration steps you want to perform from the
Self-Calibration
dialog
box.
4. Click
Next
to run the self-calibration.
NI PXIe-5667 (3.6 GHz) Getting Started Guide
|
© National Instruments
|
25