Always check the point depth of a connector with a pin depth gauge before use to determine
whether the pin depth is within the specified range.

Avoid Lateral Pressure

Never put lateral pressure on the center pin of the connector.

Always Torque Connectors to 0.9 N · m Using a Torque Wrench

Do not overtorque connectors; doing so may damage the connector center pin. Always use a
0.9 N · m (8 in-lb) torque wrench when tightening 2.92 mm, 3.5 mm, and SMA connectors.
Finger-tighten N-type connectors. 

Never

 use pliers.

Avoid Mechanical Shock

Precision connectors are designed to withstand years of normal bench handling. Handle the
connectors carefully and avoid dropping them. Mechanical shock significantly reduces their
service life.

Avoid Applying Excessive Power

The NI 5632 is rated for a specific maximum continuous input power. Exceeding the
maximum input power level permanently damages the internal components. Before making
connections, verify the maximum input levels labeled below the ports on the module front
panel.

Clean Connectors Carefully

Dirt and other contamination on the connector interfaces can affect the performance of the RF
components. To clean the connector interfaces, use only pure isopropyl alcohol as a cleaning
solvent. Do not use excessive amounts of alcohol, as this prolongs drying time. Apply the
alcohol with a foam-tipped swab instead of cotton swabs, because the cotton can become
lodged in the connector.

Use low-pressure compressed air to remove foreign particles and to dry the connector. After
cleaning, verify that the center pin has not been bent or damaged.

Store Connectors Properly

When not in use, keep the connectors covered with the dust cap included in the NI 5632
shipping kit.

Configuring the NI 5632 in MAX

Use Measurement & Automation Explorer (MAX) to configure your National Instruments
hardware. MAX informs other programs about which devices reside in the system and how
they are configured. MAX is automatically installed with NI-VNA.
1.

Launch MAX.

2.

In the Configuration pane, double-click 

Devices and Interfaces

 to see the list of installed

devices.

Installed devices appear under the name of their associated chassis.

NI PXIe-5632 Getting Started Guide

  | 

© National Instruments

  | 

11

Содержание PXIe-5632

Страница 1: ...he most recent specifications for your device Caution The protection provided by this product may be impaired if it is used in a manner not described in this document Contents Electromagnetic Compatib...

Страница 2: ...accessories Caution To ensure the specified EMC performance the length of all I O cables must be no longer than 3 m 10 ft Verifying the System Requirements To use the NI VNA instrument driver your sys...

Страница 3: ...ice kit that you need to operate the NI 5632 Your application may require additional items not included in your kit to install or operate your device Required Items A PXI Express chassis and chassis d...

Страница 4: ...Male to N Type Female 781610 01 Calibrated 5 16 Torque Wrench 8 in lbs 0 9 Nm 781615 01 Preparing the Environment Ensure that the environment you are using the NI 5632 in meets the following specifica...

Страница 5: ...ges or the metal bracket 1 Ensure the AC power source is connected to the chassis before installing the modules The AC power cord grounds the chassis and protects it from electrical damage while you i...

Страница 6: ...ched Position 9 Latch the module in place by pulling up on the ejector handle 10 Secure the device front panel to the chassis using the front panel mounting screws Note Tightening the top and bottom m...

Страница 7: ...bles that you attach to the NI 5632 to ensure that the reference plane remains constant between calibration and measurement Tip You can also use phase equal adapters after calibration to preserve the...

Страница 8: ...2 PORT 1 or PORT 2 connector ensure that there are no signal transients greater than the RF and DC specifications for the device that are being generated and sourced to the NI 5632 PORT 1 or PORT 2 co...

Страница 9: ...operation Table 3 NI 5632 Front Panel Connectors Connector Use PORT 1 Input output terminal for NI 5632 measurements SOURCE 1 IN Input terminal for the source 1 access loop SOURCE 2 IN Input terminal...

Страница 10: ...the error condition is removed Connector Care The NI 5632 is a high accuracy precision laboratory device and should be carefully handled Follow the precautions listed in this section when handling or...

Страница 11: ...ont panel Clean Connectors Carefully Dirt and other contamination on the connector interfaces can affect the performance of the RF components To clean the connector interfaces use only pure isopropyl...

Страница 12: ...ed and restart 4 Record the device identifier MAX assigns to the hardware Use this identifier when programming the NI 5632 5 Self test the device by selecting the device in the configuration tree and...

Страница 13: ...ws CVI Available at Program Files IVI Foundation IVI Drivers niVNA Microsoft Visual C C Use examples located in the NIDocDir NI VNA examples directory where NIDocDir is one of the following directorie...

Страница 14: ...can be useful for measuring device raw performance custom measurement configuration and other specialized measurements You can perform uncalibrated measurements using the NI VNA Soft Front Panel SFP...

Страница 15: ...nown verification standard Note Do not use the same calibration standards that you used for calibration to perform the verification f Save the instrument state and calibration 3 To load a saved calibr...

Страница 16: ...e chassis power rails a hardware module or the LED Caution Apply external signals only while the NI 5632 is powered on Applying external signals while the device is powered off may cause damage 1 Disc...

Страница 17: ...System Hardware Device Manager 5 If you are using a PXI controller verify that a National Instruments entry appears in the system device list and that the NI PXIe 5632 is enabled If the device is disa...

Страница 18: ...software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Ex...

Страница 19: ...your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Express...

Страница 20: ...d third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance pol...

Отзывы: