National Instruments Corporation
7
NI 5622 Calibration Procedure
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
•
Keep cables as short as possible. Long cables act as antennae, picking
up noise that can affect measurements.
•
Verify that all connections, including front panel connections, are
secure.
SMA Cable
Mini-Circuits
CBL-1.5FT-SMS
M+
,
50
Ω
, Maximum VSWR: 1.20
Frequency Range: 900 kHz to 820 MHz
Power Measurement Range:
–4 dBm to 3 dBm
Relative Shielding: –100 dB
Attenuator
Mini-Circuits
VAT-3+
Frequency Range: 900 kHz to 820 MHz
Minimum Power (W): 10 mW
Maximum VSWR: 1.20
Nominal Attenuation: 3 dB
SMA
Termination
Mini-Circuits
ANNE-50+
50
Ω
Power
Splitter
Weinschel
WA1507R
Frequency Range: 900 kHz to 820 MHz
Input Power Range: –4 dBm to 3 dBm
Maximum VSWR (Output ports): 1.15
Maximum Amplitude Tracking: 0.15 dBm
Table 3.
Required Equipment Specifications for NI 5622 External Calibration (Continued)
Required
Equipment
Recommended
Equipment
Calibration
Procedure
Specification