NI 5622 Calibration Procedure
46
ni.com
14. End the session using the niScope Close VI.
You have finished verifying the average noise density for the NI 5622.
Adjustment
An adjustment is required only once a year. Following the adjustment
procedure automatically updates the calibration date and temperature in the
digitizer EEPROM.
Complete the following steps to externally adjust the NI 5622.
1.
Make the following connections:
a.
Connect the signal generator to the source input port of the power
splitter.
b.
Connect the power meter to output port 1 of the power splitter.
c.
Connect channel 0 (IF IN) of the digitizer to output port 2 of the
power splitter.
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the following
parameter:
vi
: The instrument handle from
niScope_init
Table 8.
NI 5622 Average noise Density Published Specifications
Iteration
Range (Vpp)
Bandpass Filter
Published Specifications
(dBm/Hz)
1
0.7
NISCOPE_VAL_FALSE
–146
2
1
NISCOPE_VAL_FALSE
–143
3
1.4
NISCOPE_VAL_FALSE
–140
4
0.7
NISCOPE_VAL_TRUE
–146
5
1
NISCOPE_VAL_TRUE
–143
6
1.4
NISCOPE_VAL_TRUE
–140