background image

Troubleshooting

Why Is the ACCESS LED Off When the Chassis is On?

The LEDs may not light until the device has been configured in MAX. Before proceeding,
verify that the NI 5622 appears in MAX.

If the ACCESS LED fails to light after you power on the chassis, a problem may exist with the
chassis power rails, a hardware module, or the LED.

Caution

 

Apply external signals only while the NI 5622 is powered on. Applying

external signals while the device is powered off may cause damage.

1.

Disconnect any signals from the module front panels.

2.

Power off the chassis.

3.

Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged device.

4.

Reinstall the module in a different chassis slot.

5.

Power on the chassis.

6.

Verify that the device appears in MAX.

7.

Reset the device in MAX and perform a self-test.

Related Information

Installing the NI 5622

 on page 4

What Should I Do if the NI 5622 Doesn't Appear in
MAX?

1.

In the MAX configuration tree, expand 

Devices and Interfaces

.

2.

Expand the 

Chassis

 tree to see the list of installed devices, and press <F5> to refresh the

list.

3.

If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.

4.

Navigate to the Device Manager.

Operating System Description

 

 

Windows 8

Right-click the Start screen, and select 

All apps

»

Control Panel

»

Hardware and Sound

»

Device Manager

.

Windows 7

Select 

Start

»

Control Panel

»

Device Manager

.

Windows Vista

Select 

Start

»

Control Panel

»

System and Maintenance

»

Device

Manager

.

NI PXIe-5622 Getting Started Guide

  | 

© National Instruments

  | 

9

Содержание PXIe-5622

Страница 1: ...nt 3 Preparing the Environment 3 Installing the Software 4 Installing the NI 5622 4 Hardware Front Panel Connectors 6 Configuring the Hardware in MAX 7 Programming the NI 5622 8 Troubleshooting 9 Why...

Страница 2: ...cables must be no longer than 30 m 100 ft Verifying the System Requirements To use the NI 5622 your system must meet certain requirements For more information about minimum system requirements recomm...

Страница 3: ...n are required for installation and operation of the NI 5622 Preparing the Environment Ensure that the environment in which you are using the NI 5622 meets the following specifications Operating Envir...

Страница 4: ...Accept the prompts to complete the installation 4 When the installer completes select Restart in the dialog box that prompts you to restart shut down or restart later Installing the NI 5622 Caution T...

Страница 5: ...til it is fully inserted Figure 3 Module Installation NI PXIe 1075 2 3 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatched Position 9 Latch the module in place by pulling up on the ejecto...

Страница 6: ...ACTIVE PFI 1 IF IN CLK IN CLK OUT TTL 5 V MAX Table 1 NI 5622 Front Panel Signal Connections Connector Description Function IF IN SMA connector Analog input connection digitizes data and triggers acq...

Страница 7: ...ices If the device is still not listed power off the system ensure the device is correctly installed and restart 4 Record the device identifier MAX assigns to the hardware Use this identifier when pro...

Страница 8: ...COPE The NI SCOPE API configures and operates the device hardware and provides customizable acquisition control analysis and measurement options using LabVIEW VIs or LabWindows CVI functions C or LabW...

Страница 9: ...rent chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform a self test Related Information Installing the NI 5622 on page 4 What Should I Do...

Страница 10: ...dule in a different slot 5 Power on the chassis 6 Perform the self test again Thermal Shutdown Error If you receive an over temperature or thermal shutdown error and your device shuts down complete th...

Страница 11: ...Help Worldwide Support and Services The National Instruments website is your complete resource for technical support At ni com support you have access to everything from troubleshooting and applicati...

Страница 12: ...mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the appropriate location Help Patents in your softwa...

Отзывы: