NI PXIe-5611 Calibration Procedure
Figure 1 shows the programming flow for a complete calibration.
Figure 1.
Complete Calibration Programming Flow
Document
Post-Adjustment
Results
Document
Pre-Adjustment
Results
Calibration/
Verification
Complete
Yes
No
Verify
Meets
Test Limits?
Review Verification/
Adjustment Procedure
or Return Device
Adjust (Calibration Dates
and Temperatures Updated)
Verify