Table 6. Passband Flatness Accuracy Verification (Continued)
Amplitude Range
(V
pk-pk
)
Frequency (MHz)
As-Found Test
Limit (dB)
As-Left Test Limit
(dB)
5
1.040
0.05
—
—
6
1
±0.40
±0.25
7
20
±0.40
±0.11
8
40
±0.45
±0.10
11
0.956
0.05
—
—
12
1
±0.40
±0.25
13
20
±0.40
±0.11
14
40
±0.45
±0.10
17
0.867
0.05
—
—
18
1
±0.40
±0.25
19
20
±0.40
±0.11
20
40
±0.45
±0.10
23
0.780
0.05
—
—
24
1
±0.40
±0.25
25
20
±0.40
±0.11
26
40
±0.45
±0.10
29
0.321
0.05
—
—
30
1
±0.40
±0.25
31
20
±0.40
±0.11
32
40
±0.45
±0.10
35
0.099
0.05
—
—
36
1
±0.40
±0.25
37
20
±0.40
±0.11
38
40
±0.45
±0.10
1.
Connect the SMA (m)-to-N (f) adapter to the power sensor.
12
|
ni.com
|
PXIe-5423 Calibration Procedure